Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Stress Generation during Pecking Motion of Rotary Nickel-titanium Instruments with Different Pecking Depth
Jung-Hong Ha1, Sang Won Kwak2, Asgeir Sigurdsson3
1Department of Conservative Dentistry, School of Dentistry, IHBR, Kyungpook National University, Daegu, Korea.
Introduction:
The aim of this study was to evaluate the effect of different pecking depth on the stress generated by the screw-in forces of a rotating endodontic file in simulated canals.
Methods:
Twenty simulated resin blocks with a J-shaped curvature were used. Twenty OneG files (MicroMega, Besançon, France) were assigned for a screw-in test depending on the pecking depth in 2 groups (n = 10). The files were operated at 300 rpm, and the up and down speed was controlled at 1 mm/s stroke velocity and a 10-millisecond dwell time using a customized device. The distances (pecking depth) for the pecking motion were 2 mm or 4 mm for each group; "6 mm forward and 4 mm backward" and "6 mm forward and 2 mm backward" movements were applied, respectively, for the 2 pecking groups. During the operation, the positive and negative apical loads were recorded at a rate of 50 Hz using customized software attached to the device. The maximum negative apical load (screw-in force [SF]) was recorded, and the total energy during pecking motion until the file reached the working length (cumulative screw-in forces [CSFs]) was computed. The data were analyzed using an independent t test at a significance level of 95%.
Results:
No significant difference in SF was found between the 2 groups of pecking depths. However, the longer pecking depth (4-mm group) showed a significantly larger CSF compared with the shorter pecking depth group (P < .05).
Conclusions:
The shorter pecking depth may generate lower overall stresses for the root dentin as well as the instrument.