Factors which influence concordance among measurements obtained by different pulse oximeters currently used in some
Candelaria de la Merced Díaz-González1, Milagros de la Rosa-Hormiga2, Josefa M Ramal-López2
1Hospital Universitario Insular de Gran Canaria, Las Palmas, Islas Canarias, Spain.
Aims And Objectives:
To evaluate both concordance among those measurements obtained by three different pulse oximeters currently used by nursing professionals in hospital units and the factors which can influence this concordance.
Background:
Many models of wireless pulse oximeters in the present market do not offer possibility of calibration and, therefore, they do not ensure patients' safety in daily clinical practice.
Design:
This is a descriptive and cross-sectional study. The sample of patients (140) was selected from such hospital units, and all of them had to fulfil inclusion and exclusion criteria for participation.
Methods:
The instruments used to carry out this research were a monitor (calibrated), two models of portable wireless pulse oximeters (used for 3 years by the nursing staff, without being calibrated), a tympanic thermometer and a weather meter. The concordance correlation coefficient (CCC) was used to establish the concordance, whereas the Landis-Koch criteria were used to interpret the results.
Results:
The concordance among the three devices was considered as "good" (CCC: 0.925 and 0.974 (95%)). The CCC (0.925) for saturation measures was regarded as "very good"/"almost perfect," and the pulse measure was considered as "very good" CCC 0.974, providing in both cases a high level of concordance (CCC > 0.91).
Conclusions:
The overall concordance as regards pulse and oxygen saturation among the three pulse oximeters analysed is considered as "very good" according to the Landis-Koch criteria.
Relevance To Clinical Practice:
There exists a "very good" concordance among two wireless oximeters which have been used by the health staff for 3 years and which did not offer any possibility of calibration and one monitor which belongs to the healthcare institution. This aspect is of crucial importance in daily clinical practice, and it is also relevant to ensure patient's safety.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.