Related Experiment Video
Updated: Feb 23, 2026

E-Patient Counseling Trial E-PACO: Computer Based Education versus Nurse Counseling for Patients to Prepare for Colonoscopy
Published on: August 1, 2019
Determining the sample size required to establish whether a medical device is non-inferior to an external benchmark
Adrian Sayers1,2, Michael J Crowther3, Andrew Judge4,5
1Muscloskeletal Research Unit, School of Clinical Sciences, University of Bristol, Bristol, UK.
Objectives:
The use of benchmarks to assess the performance of implants such as those used in arthroplasty surgery is a widespread practice. It provides surgeons, patients and regulatory authorities with the reassurance that implants used are safe and effective. However, it is not currently clear how or how many implants should be statistically compared with a benchmark to assess whether or not that implant is superior, equivalent, non-inferior or inferior to the performance benchmark of interest.We aim to describe the methods and sample size required to conduct a one-sample non-inferiority study of a medical device for the purposes of benchmarking.
Design:
Simulation study.
Setting:
Simulation study of a national register of medical devices.
Methods:
We simulated data, with and without a non-informative competing risk, to represent an arthroplasty population and describe three methods of analysis (z-test, 1-Kaplan-Meier and competing risks) commonly used in surgical research.
Primary Outcome:
We evaluate the performance of each method using power, bias, root-mean-square error, coverage and CI width.
Results:
1-Kaplan-Meier provides an unbiased estimate of implant net failure, which can be used to assess if a surgical device is non-inferior to an external benchmark. Small non-inferiority margins require significantly more individuals to be at risk compared with current benchmarking standards.
Conclusion:
A non-inferiority testing paradigm provides a useful framework for determining if an implant meets the required performance defined by an external benchmark. Current contemporary benchmarking standards have limited power to detect non-inferiority, and substantially larger samples sizes, in excess of 3200 procedures, are required to achieve a power greater than 60%. It is clear when benchmarking implant performance, net failure estimated using 1-KM is preferential to crude failure estimated by competing risk models.
More Related Videos
04:53A Clinical Trial Assessing the Safety, Efficacy, and Delivery of Olive-Oil-Based Three-Chamber Bags for Parenteral Nutrition
Published on: September 20, 2019
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Bioequivalence Data: Statistical Interpretation
Sample Size Calculation
The sample size for the given experiment or sampling effort is fundamental to any study design. Sample size decides the number of...
One-Way ANOVA: Unequal Sample Sizes
Bonferroni Test
The means of different samples are first paired in all possible combinations.
The null hypothesis of the...
One-Way ANOVA: Equal Sample Sizes
Different sample means can result in different values for the variance estimate: variance between samples. This is because the variance between samples is calculated as the product of the sample size and the variance between the...