Construction and Test Results of Coil 2 of a Three-Coil 800-MHz REBCO Insert for the 1.3-GHz High-Resolution NMR
Juan Bascuñán1, Philip Michael2, Seungyong Hahn3
1MIT Francis Bitter Magnet Laboratory/PSFC, Magnet Technology Division, Cambridge, MA 02139 USA.
Abstract:
This paper focuses on the construction and test results of Coil 2 that is part of a trio of nested coils composing the REBCO 800 MHz insert. Upon its completion, the REBCO 800 MHz insert will be placed in the bore of a 500 MHz low temperature superconducting (LTS) NMR magnet (L500) to form the MIT 1.3 GHz high-resolution NMR magnet. Coil 2 is a stack of 32 double pancake (DP) coils wound with 6-mm wide REBCO tape using the no-insulation (NI) technique. Each pancake is wound on a stainless steel inner supporting ring to prevent the collapsing of its crossover due to the external pressure exerted by the winding pack. Coil 2 will be constructed in the following sequence: 1) after winding each DP will be individually tested in a bath of liquid nitrogen at atmospheric pressure to determine its current carrying capabilities; 2) DPs will be then assembled as a stack with interconnecting joints, and 3) as in Coil 1, each pancake will be overbanded with a stainless steel tape, this time to a thickness of 5 mm, thickness determined by a stress analysis previously performed. Finally the fully assembled Coil 2 will be tested in liquid nitrogen at 77 K and then in liquid helium at 4.2 K. We present here details of the stress analysis leading to the sizing of the DP inner supporting stainless steel ring and of the overbanding thickness required. Test results include coil index, critical current, charging time constant.
More Related Videos
08:42High-Sensitivity Nuclear Magnetic Resonance at Giga-Pascal Pressures: A New Tool for Probing Electronic and Chemical Properties of Condensed Matter under Extreme Conditions
Published on: October 10, 2014
05:26Author Spotlight: Low-Cost Electroencephalographic Recording System Combined with a Millimeter-Sized Coil to Transcranially Stimulate the Mouse Brain In Vivo
Published on: May 26, 2023
Related Concept Videos
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Atomic Nuclei: Magnetic Resonance
NMR Spectroscopy: Spin–Spin Coupling
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...
2D NMR: Overview of Heteronuclear Correlation Techniques
NMR Spectrometers: Overview