Optimization of the Monte Carlo simulation model of NaI(Tl) detector by Geant4 code
Hoang Duc Tam1, Nguyen Thi Hai Yen1, Le Bao Tran2
1Faculty of Physics, Ho Chi Minh City University of Pedagogy, Ho Chi Minh City, Vietnam.
Optimizing the geometrical parameters of sodium iodide (NaI(Tl)) detectors using Geant4 simulations improves accuracy. This method ensures reliable Monte Carlo simulations for NaI(Tl) detector applications.
Area of Science:
- Nuclear Instrumentation
- Computational Physics
- Radiation Detection
Background:
- Accurate modeling of sodium iodide (NaI(Tl)) detectors is crucial for reliable radiation measurements.
- Geometrical parameters from manufacturers may differ from actual detector dimensions, impacting simulation fidelity.
- Monte Carlo simulations, particularly with Geant4, are widely used but sensitive to input parameters.
Purpose of the Study:
- To optimize the geometrical parameters of a NaI(Tl) detector model within the Geant4 simulation framework.
- To develop a systematic procedure for refining detector geometry in simulations.
- To minimize discrepancies between simulated and experimental results for NaI(Tl) detectors.
Main Methods:
- Utilized Geant4, a Monte Carlo toolkit, for simulating NaI(Tl) detector response.
- Developed and applied a procedure to optimize detector geometrical parameters.
- Compared simulation outputs with experimental data for validation.
Main Results:
- Achieved good agreement between simulated and experimental values for the NaI(Tl) detector.
- The optimized model demonstrated a maximum discrepancy of 3.96% for full-energy peak efficiency.
- Energy resolution and peak-to-Compton ratio showed discrepancies of 1.69% and 3.50%, respectively.
Conclusions:
- The proposed optimization procedure effectively refines NaI(Tl) detector geometrical parameters in Geant4 simulations.
- Accurate geometrical parameter optimization leads to significantly improved simulation accuracy.
- This method enhances the reliability of Monte Carlo simulations for NaI(Tl) detector characterization.