Positron Emission Tomography Using 64-Copper as a Tracer for the Study of Copper-Related Disorders
Published on: April 28, 2023
Assessing the absolute quantitative accuracy of Positron Emission Tomography for Cu-64 using traceable calibrated
B E Zimmerman1, D E Bergeron1, J P Edgerton2
1Physical Measurement Laboratory, National Institute of Standards and Technology, Gaithersburg, MD 20899-8462, USA.
Abstract:
Using uniform cylindrical phantoms containing calibrated solutions of 18F and 64Cu, we evaluated for the first time the accuracy with which the activity concentration of 64Cu can be quantified on an absolute basis using Positron Emission Tomography (with X-ray Computed Tomography, PET-CT). The scanner was first calibrated for 18F using the manufacturer's calibration protocol and a phantom with an activity concentration value traceable to the U.S. National standard. By using a similarly calibrated 18F solution phantom, we were able to determine a correction factor that can be applied to the 64Cu imaging data that gave a result that is consistent with 100% recovery with a combined standard uncertainty of 2%. We also demonstrate how a calibrated, solid phantom containing 68Ge as a long-lived 18F surrogate can be used to monitor and transfer the correction factor to other studies.
More Related Videos
14:19A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
Published on: February 1, 2016
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET