Related Experiment Video
Updated: Feb 20, 2026

Author Spotlight: A Non-Intubated Video-Assisted Thoracoscopic Surgery with Multimodal Analgesia and Sevoflurane Inhalation Anesthesia
Published on: May 26, 2023
Digital Devices Improve Chest Tube Management: Results from a Prospective Randomized Trial
Marco Chiappetta1, Filippo Lococo2, Dania Nachira1
1Unità operativa di Chirurgia Toracica, Università Cattolica del Sacro Cuore, Fondazione Policlinico Universitario A. Gemelli, Roma, Italy.
Background:
Digital devices could help clinical practice measuring the air leak, but their role is still debated. Aim of this study is to test advantages using these devices.
Methods:
From June 2012 to May 2015, we enrolled 95 patients undergoing lobectomy or wedge resection in a prospective randomized trial. Patients were divided into two groups: group D (digital, 50 patients) evaluated with digital device and group E (empirical, 45 patients) evaluated with water seal. Logistic regression analysis and receiver operating characteristic (ROC) curve analysis were used to select and assess predictors of persistent air leak. In group D, chest drain was removed if the air leak was absent or < 0.5 L/h for 12 consecutive hours and in group E if clinical signs of air leak were absent.
Results:
Mean hospitalization and mean chest tube stay was 6.0 ± 3.3 and 4.4 ± 3.2 days, respectively, prolonged air leak occurred in nine (8%). Hospitalization was 5.8 ± 2.5 versus 6.2 ± 4.2 days (p = 0.5), while mean chest tube stay was 4.1 ± 2.0 versus 4.6 ± 3.5 days (p = 0.4) in group D and group E, respectively. Clamping test was needed in one patient in group D and in seven in group E (p = 0.019). At multivariate analysis, heart disease (p < 0.0001), lobectomy (p < 0.0001), fused fissure (p < 0.0001), and air leakage in first postoperative (p.o.) day were predictors of persistent air leak (AUC on the ROC curve of 69.7%, sensibility: 77.8%). In group D, an air leak value > 0.2 L/h with spikes over 0.5 L/h in third p.o. was predictive of persistent air leak, with chest tube duration of 7.73 ± 5.20 versus 4.32 ± 1.33 days (AUC: 83%, sensitivity: 80%, p < 0.0001).
Conclusion:
In our experience, digital devices reduced observer variability and mistakes in chest tube management, thus identifying patients at risk for prolonged air leak.
Related Concept Videos
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History:
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...