Simultaneous determination of renal function biomarkers in urine using a validated paper-based microfluidic
Eduardo Luiz Rossini1, Maria Izabel Milani1, Emanuel Carrilho2
1Instituto de Química, Universidade Estadual Paulista "Julio de Mesquita Filho"- UNESP, R. Prof. Francisco Degni 55, P.O. Box 355, 14801-970, Araraquara, SP, Brazil.
This study presents a low-cost paper device for simultaneously measuring creatinine and uric acid in urine. The validated microfluidic analytical device offers a simple, colorimetric method for assessing renal function biomarkers.
Area of Science:
- Analytical Chemistry
- Biomarker Detection
- Microfluidics
Background:
- Renal function is commonly assessed by measuring urine biomarkers like creatinine and uric acid.
- Accurate and accessible quantification methods are crucial for effective diagnosis and monitoring.
- Existing methods can be complex or expensive, limiting widespread use.
Purpose of the Study:
- To develop and validate a paper-based microfluidic analytical device (μPAD) for simultaneous quantification of creatinine (CRN) and uric acid (UA) in urine.
- To provide an inexpensive, simple, and user-friendly colorimetric method for renal function biomarker analysis.
- To optimize the detection method using chemometrics.
Main Methods:
- Fabrication of a microfluidic paper device with distinct testing and uptake zones.
- Colorimetric detection of creatinine via the Jaffé reaction.
- Colorimetric detection of uric acid via Fe³⁺ reduction and 1,10-phenanthroline complexation.
- Optimization of reaction conditions using chemometrics.
Main Results:
- The μPAD demonstrated good linear correlations between analyte concentrations and effective intensities.
- Satisfactory repeatability was achieved for both creatinine and uric acid measurements.
- The limits of detection were 15.7 mg/L for CRN and 16.5 mg/L for UA, with linear ranges of 50-600 mg/L and 50-500 mg/L, respectively.
- No statistically significant differences were observed compared to a chromatographic method.
Conclusions:
- The validated μPAD is a reliable and cost-effective tool for simultaneous quantification of urinary creatinine and uric acid.
- This technology offers a promising alternative for point-of-care or resource-limited settings for renal function assessment.
- The developed method is simple, easy to use, and provides accurate results comparable to established techniques.
Related Concept Videos
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Determination of Renal Drug Clearance: Graphical and Midpoint Methods
The graphical method involves plotting the rate of drug excretion in urine against the plasma drug concentration. By analyzing the graph, the clearance can be calculated and obtained. Drugs rapidly excreted by the kidneys exhibit a...
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...