Pulsed electron avalanche knife (PEAK) PlasmaBlade™ in pacemaker and defibrillator procedures
Elif Kaya1, Matthias Totzeck2, Tienush Rassaf2
1Department of Cardiology and Vascular Medicine, West German Heart and Vascular Center Essen, University Hospital Essen, Hufelandstrasse 55, 45147, Essen, Germany. Elif.Kaya@uk-essen.de.
Background:
The pulsed electron avalanche knife (PEAK) PlasmaBlade™ is an innovative electrosurgical device that uses a novel technology to cut tissues. It has been proven to be safe and feasible in ear, nose, and throat surgery, but there are only limited data concerning the use of PlasmaBlade™ instead of conventional electrocautery in cardiac implantable electronic device (CIED) procedures except for generator replacements.
Methods:
We conducted a retrospective, single-center study with patients undergoing CIED surgery at our center between December 2015 and March 2017 and evaluate the feasibility and the clinical outcome of the PlasmaBlade™.
Results:
282 patients (mean age 70.7 ± 12.9 years; 65.6% male) were included, of which 119 (42.2%) underwent pacemaker implantation, 95 (33.7%) implantable cardioverter defibrillator implantation, and 68 (24.1%) received a generator replacement. At the time of the procedure, 55 patients (19.5%) were on dual antiplatelet therapy, and 109 (38.7%) patients were on oral anticoagulation (30.5% vitamin K antagonists, 8.2% novel oral anticoagulants). The overall perioperative complication rate was 3.9%. Device-pocket hematoma occurred in 9 patients (3.2%) requiring further surgery. No lead damage was seen within a follow-up of 6 months. One patient presented with device-pocket infection 2.9 months after implantation of a cardiac resynchronization therapy defibrillator requiring CIED system extraction.
Conclusions:
Replacing conventional electrocautery by PlasmaBlade™ for CIED procedures is feasible with a moderate rate of perioperative complications compared to the literature. Studies comparing the PlasmaBlade™ with conventional electrocautery are necessary to investigate whether PlasmaBlade™ offers an additional benefit over conventional electrocautery.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Dysrhythmias VI: Management of Dysrhythmias
Cardiopulmonary Resuscitation IV: Pharmacological Management