A new approach for accurate mass assignment on a multi-turn time-of-flight mass spectrometer
Toshinobu Hondo1,2, Kirk R Jensen2, Jun Aoki2
11 MS-Cheminformatics, LLC, Inabe-gun, Japan.
Abstract:
A simple, effective accurate mass assignment procedure for a time-of-flight mass spectrometer is desirable. External mass calibration using a mass calibration standard together with an internal mass reference (lock mass) is a common technique for mass assignment, however, using polynomial fitting can result in mass-dependent errors. By using the multi-turn time-of-flight mass spectrometer infiTOF-UHV, we were able to obtain multiple time-of-flight data from an ion monitored under several different numbers of laps that was then used to calculate a mass calibration equation. We have developed a data acquisition system that simultaneously monitors spectra at several different lap conditions with on-the-fly centroid determination and scan law estimation, which is a function of acceleration voltage, flight path, and instrumental time delay. Less than 0.9 mDa mass errors were observed for assigned mass to charge ratios ( m/z) ranging between 4 and 134 using only 40Ar+ as a reference. It was also observed that estimating the scan law on-the-fly provides excellent mass drift compensation.
Related Concept Videos
Tandem Mass Spectrometry
Mass Analyzers: Common Types
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
MALDI-TOF Mass Spectrometry
Mass Spectrum: Interpretation