Experimental confirmation of the atomic force microscope cantilever stiffness tilt correction
1Nanomechanical Properties Group, National Institute of Standards and Technology, 100 Bureau Drive, STOP 8370, Gaithersburg, Maryland 20899, USA.
Abstract:
The tilt angle (angle of repose) of an AFM cantilever relative to the surface it is interrogating affects the effective stiffness of the cantilever as it analyzes the surface. For typical AFMs and cantilevers that incline from 10° to 15° tilt, this is thought to be a 3%-7% stiffness increase correction. While the theoretical geometric analysis of this effect may have reached a consensus that it varies with cos-2 θ, there is very little experimental evidence to confirm this using AFM cantilevers. Recently, the laser Doppler vibrometry thermal calibration method utilized at NIST has demonstrated sufficient stiffness calibration accuracy, and precision to allow a definitive experimental confirmation of the particular trigonometric form of this tilt effect using a commercial microfabricated AFM cantilever specially modified to allow strongly tilted (up to 15°) effective cantilever stiffness measurements.