Survey of 5 mm small-field output factor measurements in Australia
Christopher P Oliver1, Duncan J Butler1, Viliami Takau1
1Australian Radiation Protection and Nuclear Safety Agency, Yallambie, Vic, Australia.
Abstract:
The Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) held a comparison exercise in April 2016 where participants came to ARPANSA and measured the output factor of a nominal 5 mm cone attached to the ARPANSA Elekta Synergy (Elekta, Crawley, UK) linear accelerator. The goal of the exercise was to compare the consistency and methods used by independent medical physicists in measuring small-field output factors. ARPANSA provided a three-dimensional scanning tank for detector setup and positioning, but the participants were required to measure the output factor with their own detectors. No information regarding output factors previously measured was supplied to participants to make each result as independent as possible. Fifteen groups travelled to ARPANSA bringing a wide range of detectors and methods. A total of 30 measurements of the output factor were made. The standard deviation of the measurements (excluding one expected outlier from an uncorrected ionization chamber measurement) was 3.6%. The results provide an insight into the consistency of small-field dosimetry being performed in Australia and New Zealand at the present time.
Related Concept Videos
Surveys
Introduction to Surveying, Plane Surveying and Geodetic Surveys
Voltammetry: Factors Affecting Measurements
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Errors and Mistakes in Surveying
Types of Surveys