The Short Cosyntropin Test Revisited: New Normal Reference Range Using LC-MS/MS.
Grethe Å Ueland1,2,3, Paal Methlie1,2,3, Marianne Øksnes1,2,3
1Department of Clinical Science, University of Bergen, Bergen, Norway.
New liquid chromatography tandem mass spectrometry (LC-MS/MS) cutoffs for the cosyntropin test improve diagnosis of adrenal insufficiency (AI) and nonclassical congenital adrenal hyperplasia (NCCAH). These precise LC-MS/MS ranges reduce false positives and negatives in steroid testing.
Area of Science:
- Endocrinology
- Clinical Chemistry
- Diagnostic Medicine
Background:
- Current cosyntropin test cutoffs for adrenal insufficiency (AI) and nonclassical congenital adrenal hyperplasia (NCCAH) rely on nonstandardized immunoassays.
- Liquid chromatography tandem mass spectrometry (LC-MS/MS) offers direct steroid measurement, necessitating updated reference ranges.
Purpose of the Study:
- To establish serum cortisol and 17-hydroxyprogesterone (17-OHP) cutoff values using LC-MS/MS after cosyntropin stimulation.
- To compare LC-MS/MS results with those from routine immunoassays.
Main Methods:
- Cosyntropin testing performed on 138 healthy volunteers and 94 patients.
- Steroids assayed by LC-MS/MS and compared with two common immunoassays.
- Cutoff levels determined using percentile calculations in healthy subjects.
Main Results:
- LC-MS/MS cortisol cutoffs were 412 and 485 nmol/L at 30 and 60 minutes, respectively.
- New cutoffs reclassified 22% of subjects with AI to normal.
- LC-MS/MS 17-OHP cutoff levels were 8.9 and 9.0 nmol/L at 30 and 60 minutes.
Conclusions:
- LC-MS/MS provides lower, more accurate cutoff levels for cortisol and 17-OHP post-cosyntropin stimulation.
- Elimination of immunoassay cross-reactivity enhances diagnostic precision.
- This method reduces false-positive AI diagnoses and false-negative NCCAH diagnoses.
Related Concept Videos
Variation: Normal Distribution, Range, and Standard Deviation
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
Inertial Frames of Reference
Non-inertial Frames of Reference
Normal Distribution
Normal Stress
When a rod is under axial loading, the internal forces and corresponding stress are normal to the plane of the section, so it is termed normal stress. It's important to...