A new minimally invasive technique for lead revision of perc-paddle leads
Olivier De Coster1,2, Patrice Forget3, Maarten Moens4
1Pain Clinic, Algemeen Ziekenhuis Delta (AZ Delta), Roeselare, Belgium.
Abstract:
To report on a less-invasive technique for replacing a broken lead in a spinal cord stimulation (SCS) device that makes use of St. Jude Medical's "Epiducer" device. A 53-year-old woman suffered a loss of stimulation on her internal pulse generator (IPG), which was found to have a broken lead. The broken lead was withdrawn using the Epiducer device with minimal invasiveness and without pain. A new lead was put in place, and successful stimulation using the IPG resumed. Follow-up 3 weeks later showed a well-functioning SCS system. An Epiducer can be used to revise the SCS system in a minimal invasive way without a new puncture. Follow-up study of this technique as well as others for revising an SCS system should be encouraged and used in comparison study.
Related Concept Videos
Phase-lead and Phase-lag Controllers
Time and frequency -Domain Interpretation of Phase-lead Control
The design of phase-lead control involves the strategic placement of poles and zeros to balance steady-state error and system...
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Scientific Laws and Theories
Overview of Microscopy Techniques