Impact of Cone Beam Computed Tomography Dose in Pre-Surgical Implant Analysis
Nabil El Sahili1, Ibrahim Nasseh2, Antoine Berberi3
1School of Dentistry, Lebanese University, Beirut, Lebanon.
Objectives:
Cone-Beam Computed Tomography (CBCT) produces vital information required for the accurate and prudent placement of dental implants. Lack of standardization between CBCT machines may result in unsafe patient exposure to harmful radiation; higher doses are not necessarily associated with improved image quality.
Aim:
The study aimed to assess the influence of low- and high-dose milliamperage settings on CBCT images for objective and subjective implant planning.
Methods:
Two dry skulls (4 hemi-maxillary segments of the maxilla and 4 hemi-maxillary segments of the mandible) were scanned under low (2 mA) and high (6.3 mA) dosage settings using a CBCT (Carestream CS 9300). Cross-sectional slices of both image qualities were evaluated by five expert clinicians, for image quality for implant planning and objective bone measurements.
Results:
There were no significant differences in bone measurements taken on high or low dose images (p > 0.05). In qualitative image assessments, assessment and image quality for almost all observers were independent of each other. For planning posterior mandibular implant placement, increased dosage improved concordance and kappa values between low and high dose images.
Conclusion:
Reduction in milliamperage did not affect diagnostic image quality for objective bone measurements and produced sufficient intra-rater reliability for qualitative assessment; therefore dose reduction can be achieved without compromising diagnostic decision- making.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...
Imaging Studies III: Computed Tomography
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
pre-mRNA Processing
Once about 20-40 ribonucleotides have been joined together by RNA polymerase, a group of enzymes adds a “cap” to the 5’ end of the growing transcript. In this process, a 5’ phosphate is replaced by modified guanosine that has a methyl group attached to it (7-Methyl...
Impact of Groups on Groups