Anterior segment ultrasound biomicroscopy image analysis using ImageJ software: Intra-observer repeatability and
Azam Qureshi1, Haoxing Chen2, Osamah Saeedi3,4
1University of Maryland School of Medicine, Baltimore, MD, USA. azam.qureshi@som.umaryland.edu.
Purpose:
In this novel study, we demonstrate a standardized imaging and measurement protocol of anterior segment (AS) structures with reliability analysis using ultrasound biomicroscopy (UBM) and ImageJ software.
Methods:
Ten pediatric and young adult patients undergoing examination under anesthesia for AS pathology were imaged using UBM. Four trained observers analyzed 20 images using ImageJ. Forty-five structural parameters were measured. Those that relied on the trabecular-iris angle (TIA) as a reference landmark were labeled TIA-dependent (TD) and all others were labeled non-TIA dependent (NTD). Intra-observer repeatability (IOR) and inter-observer agreement (IOA) of measurements were determined using coefficient of variation (CV) and intra-class correlation (ICC) followed by assessment of Bland-Altman plots (BAP) for each pair of observers, respectively.
Results:
For NTD parameters, non-ciliary body (CB) related measurements showed CV range 0.60-16.22% and ICC range 0.84-0.89, whereas CB-related parameters showed CV range 2.86-23.40% and ICC range 0.29-0.92. For TD parameters, parameters < 2 degrees removed from reference showed CV range 0.02-5.40% and ICC range 0.89-1.00, whereas parameters > 1 degree removed showed CV range 0.63-27.44% and ICC range 0.22-1.00. No systematic proportional bias was detected by BAPs.
Conclusions:
Preplaced landmarks yielded good IOR and IOA in quantitative assessment of AS structures that were NTD and non-CB-related or less removed from the reference. CB-related NTD measurements varied greatly in IOR and IOA, indicating protocol modifications or CB qualitative assessments needed to improve accuracy. Variability in TD measurements increased the further removed from the reference, which supports implementation of a reliable reference landmark to minimize variation.
More Related Videos
09:36Outer-Boundary Assisted Segmentation and Quantification of Trabecular Bones by an Imagej Plugin
Published on: March 14, 2018
10:38Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
Published on: January 16, 2019
Related Concept Videos
Naturalistic Observations
Observational Learning
Observational Studies
There are three types of observational studies – Prospective, retrospective, and cross-sectional.
Prospective Study
Prospective studies, also known as longitudinal or cohort studies, are carried out by collecting future data from groups sharing similar characteristics. One...
Actor-Observer Effect
Data Collection by Observations
An astronomer viewing the motion and brightness of stars in the sky and recording the data is an example of observational data collection. A botanist recording...
Statistical Software for Data Analysis and Clinical Trials