Cone-beam CT-based inter-fraction localization errors for tumors in the pelvic region
Cristina Garibaldi1, Cristiana Fodor2, Giulia Riva3
1Radiation Research Unit, European Institute of Oncology, Milano, Italy.
Purpose:
To evaluate inter-fraction tumor localization errors (TE) in the RapidArc® treatment of pelvic cancers based on CBCT. Appropriate CTV-to PTV margins in a non-IGRT scenario have been proposed.
Methods:
Data of 928 patients with prostate, gynecological, and rectum/anal canal cancers were retrospectively analyzed to determine systematic and random localization errors. Two protocols were used: daily online IGRT (d-IGRT) and weekly IGRT. The latter consisted in acquiring a CBCT for the first 3 fractions and subsequently once a week. TE for patients who underwent d-IGRT protocol were calculated using either all CBCTs or the first 3.
Results:
The systematic (and random) TE in the AP, LL, and SI direction were: for prostate bed 2.7(3.2), 2.3(2.8) and 1.9(2.2) mm; for prostate 4.2(3.1), 2.9(2.8) and 2.3(2.2) mm; for gynecological 3.0(3.6), 2.4(2.7) and 2.3(2.5) mm; for rectum 2.8(2.8), 2.4(2.8) and 2.3(2.5) mm; for anal canal 3.1(3.3), 2.1(2.5) and 2.2(2.7) mm. CTV-to-PTV margins determined from all CBCTs were 14 mm in the AP, 10 mm in the LL and 9-9.5 mm in the SI directions for the prostate and the gynecological groups and 9.5-10.5 mm in AP, 9 mm in LL and 8-10 mm in the SI direction for the prostate bed and the rectum/anal canal groups. If assessed on the basis of the first 3 CBCTs, the calculated CTV-to-PTV margins were slightly larger.
Conclusions:
without IGRT, large CTV-to-PTV margins up to 15 mm are required to account for inter-fraction tumor localization errors. Daily IGRT should be used for all hypo-fractionated treatments to reduce margins and avoid increased toxicity to critical organs.
Related Concept Videos
Local Anesthetics: Clinical Application as Intravenous Regional Anesthesia
One of the advantages of...
Fundamental Attribution Error
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Random Error
Margin of Error