Feasibility of setting up generic alert levels for maximum skin dose in fluoroscopically guided procedures
Hannu Jarvinen1, Jad Farah2, Teemu Siiskonen1
1Radiation and Nuclear Safety Authority (STUK), P.O. Box 14, 00881 Helsinki, Finland.
Purpose:
The feasibility of setting-up generic, hospital-independent dose alert levels to initiate vigilance on possible skin injuries in interventional procedures was studied for three high-dose procedures (chemoembolization (TACE) of the liver, neuro-embolization (NE) and percutaneous coronary intervention (PCI)) in 9 European countries.
Methods:
Gafchromic® films and thermoluminescent dosimeters (TLD) were used to determine the Maximum Skin Dose (MSD). Correlation of the online dose indicators (fluoroscopy time, kerma- or dose-area product (KAP or DAP) and cumulative air kerma at interventional reference point (Ka,r)) with MSD was evaluated and used to establish the alert levels corresponding to a MSD of 2 Gy and 5 Gy. The uncertainties of alert levels in terms of DAP and Ka,r, and uncertainty of MSD were calculated.
Results:
About 20-30% of all MSD values exceeded 2 Gy while only 2-6% exceeded 5 Gy. The correlations suggest that both DAP and Ka,r can be used as a dose indicator for alert levels (Pearson correlation coefficient p mostly >0.8), while fluoroscopy time is not suitable (p mostly <0.6). Generic alert levels based on DAP (Gy cm2) were suggested for MSD of both 2 Gy and 5 Gy (for 5 Gy: TACE 750, PCI 250 and NE 400). The suggested levels are close to the lowest values published in several other studies. The uncertainty of the MSD was estimated to be around 10-15% and of hospital-specific skin dose alert levels about 20-30% (with coverage factor k = 1).
Conclusions:
The generic alert levels are feasible for some cases but should be used with caution, only as the first approximation, while hospital-specific alert levels are preferred as the final approach.
Related Concept Videos
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Maximum Deflection
The maximum deflection occurs at a specific point, known as point O, where the tangent to the deflection curve is horizontal. To find point O, the slope of the tangent at any...
Maximum Power Transfer
By substituting the entire circuit with...
Maximum Size of Aggregate
The Maximum Power Transfer Theorem
The load connected draws the current, and the circuit delivers the power to the load. The alternating current flowing through the load is determined using the rectangular form of voltages, currents, network impedance, and load impedance. The average power delivered to the load is obtained from the product of the square of current and load resistance.
Maximum Power Flow and Line Loadability