Related Experiment Video
Updated: Feb 13, 2026

The Portable Chemical Sterilizer PCS, D-FENS, and D-FEND ALL: Novel Chlorine Dioxide Decontamination Technologies for the Military
Published on: June 29, 2014
Pupillometer Use: Validation for Use in Military and Occupational Medical Surveillance and Response to
Luke Mease1,2, Reema Sikka3,4, Randall Rhees1,5
1US Army Health Clinic-Dugway Proving Ground, 5116 Kister Ave, Dugway, UT.
Introduction:
To analyze the effectiveness and suitability of pupillometer use in military and occupational medicine, specifically when pupil size is measured as part of medical surveillance. Pupil size is the most sensitive physical exam finding in vapor exposure to substances that inhibit acetylcholinesterase, such as nerve agent (chemical warfare) and organophosphates (used in agriculture). Pupillometer use permits real-time, accurate pupil measurements, which are of significant value in occupational setting where exposure to organophosphates is suspected and in dynamic military settings where it may be unclear if service members were exposed to nerve agent or not.
Materials And Methods:
In a worker population enrolled in medical surveillance including pupil size measurement, pupils were measured using a Colvard pupillometer, whereas their pupil size had previously been measured by manual measurement. Pupil size was compared pre- and post-pupillometer implementation.
Results:
Pupil size range was broader post-pupillometer implementation, reflecting accepted ranges of physiologically normal pupil sizes. The correlation between pupil sizes pre- and post-pupillometer was low, and the overall mean pupil diameters were statistically different between the two modalities (p < 0.0001), strongly suggesting that pupillometer use helped correct inaccurate pupil size estimates. In two real-world situations, pupillometer use proved helpful in evaluating workers who may have been exposed to an organophosphate, providing information that was immediately available and clinically relevant as part of the initial medical evaluation/physical exam.
Conclusions:
Pupillometer use is feasible and robustly increases the precision and accuracy of pupil size measurement in military and occupational medicine settings caring for workers potentially exposed to organophosphates. It provides near-real-time quantitative data that are immediately pertinent to determining if a service member or employee has been exposed to these agents.
Related Concept Videos
Spasmolytic Agents: Chemical Classification
A major class of centrally acting spasmolytics is the α2-agonist, such as tizanidine. These drugs bind to α2-adrenoceptors, inhibiting the release of the excitatory neurotransmitter glutamate. They also...
Chemical Agents for Microbial Control
Pharmacokinetic–Pharmacodynamic Relationship: Exposure, Response and Effect
Reliability and Validity
Principles of Disease Surveillance
Immune Surveillance by NK Cells and Phagocytes
Natural Killer Cells: The Fast Responders
NK cells are large granular lymphocytes found in the blood and lymphatic system. These...