Gradient-echo-based 3D submillisecond echo time pulmonary MR imaging: a preliminary usability study on clinical and
Soon Ho Yoon1, Chanhee Lee2, Jinil Park2
11 Department of Radiology, Seoul National University College of Medicine, Seoul National University Hospital , Seoul , South Korea.
Objective:
To preliminarily investigate a technical feasibility of a submillisecond echo time concurrent-dephasing-and-excitation (CODE) sequence for pulmonary MRI on clinical and preclinical MR scanners Methods: CODE imaging (echo time, 0.14 ~ 0.18 ms) was performed with American College of Radiology phantom at 3 T, 7 healthy volunteers at 1.5 and 3 T, 10 rabbits at 3 T, and 2 rodents at 9.4 T. Signal-to-noise ratio was compared in phantom. Image quality of human MRI was visually assessed on a 5-point scale for comparison between CODE and conventional lung MRI sequences. Visibility of bronchi, subcentimeter nodules, and MR air-bronchogram were assessed in animal studies.
Results:
In phantom study, signal-to-noise ratio was higher with CODE than with original three-dimensional ultrashort-echo time sequence (106.71 ± 4.32 vs 91.66 ± 3.54; p < 0.001). Image quality of human MRI was better with CODE than with conventional MRI sequences (p ≤ 0.002). Bronchi remained traceable up to the fifth bronchial generation in CODE images in rabbits and rodents. 95.2% of metastatic nodules (diameter, 1.5 ± 0.4 mm) and 93.8% of MR air-bronchogram (diameter, 0.9 ± 0.2 mm) in rabbits.
Conclusion:
Submillisecond echo time pulmonary MRI was technically feasible by using CODE on various MR scanners. Advances in knowledge: CODE can be a practical alternative for lung MRI on both clinical and pre-clinical scanners, without challenges of free-induction-decay-based ultrashort-echo time sequences.
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
What is an Electrochemical Gradient?
The chemical gradient relies on differences in the abundance of a substance on the outside versus the inside of a cell and flows from areas of high to low ion concentration. In contrast, the electrical gradient revolves around an...
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Pulmonary Embolism II: Diagnostic Studies and Interprofessional Care
Hypertension III: Clinical Manifestations and Diagnostic Studies
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies