Evaluation of Ophthalmic Surgical Instrument Sterility Using Short-Cycle Sterilization for Sequential Same-Day Use
David F Chang1, Nikki Hurley2, Nick Mamalis3
1Altos Eye Physicians, private practice, Los Altos, California.
Purpose:
The common practice of short-cycle sterilization for ophthalmic surgical instrumentation has come under increased regulatory scrutiny. This study was undertaken to evaluate the efficacy of short-cycle sterilization processing for consecutive same-day cataract procedures.
Design:
Testing of specific sterilization processing methods by an independent medical device validation testing laboratory.
Participants:
Phaco handpieces from 3 separate manufacturers were tested along with appropriate biologic indicators and controls using 2 common steam sterilizers.
Methods:
A STATIM 2000 sterilizer (SciCan, Canonsburg, PA) with the STATIM metal cassette, and an AMSCO Century V116 pre-vacuum sterilizer (STERIS, Mentor, OH) using a Case Medical SteriTite container (Case Medical, South Hackensack, NJ) rigid container were tested using phaco tips and handpieces from 3 different manufacturers. Biological indicators were inoculated with highly resistant Geobacillus stearothermophilus, and each sterility verification test was performed in triplicate. Both wrapped and contained loads were tested with full dry cycles and a 7-day storage time to simulate prolonged storage. In adherence with the manufacturers' instructions for use (IFU), short cycles (3.0-3.5-minute exposure times) for unwrapped and contained loads were also tested after only 1 minute of dry time to simulate use on a consecutive case. Additional studies were performed to demonstrate whether any moisture present in the load containing phaco handpieces postprocessing was sterile and would affect the sterility of the contents after a 3-minute transit/storage time. This approximated the upper limit of time needed to transfer a containment device to the operating room.
Main Outcome Measures:
Presence or absence of microbial growth from cultured test samples.
Results:
All inoculated test samples from both sterilizers were negative for growth of the target organism whether the full dry phase was interrupted or not. Pipetted postprocessing moisture samples and swabs of the handpieces were also negative for growth after a 3-minute transit/storage time.
Conclusions:
These studies support the use of unwrapped, short-cycle sterilization that adheres to the IFU of these 2 popular Food and Drug Administration-cleared sterilizers for sequential same-day cataract surgeries. A full drying phase is not necessary when the instruments are kept within the covered sterilizer containment device for prompt use on a sequential case.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Ophthalmic Drug Delivery Systems
The Carbon Cycle
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...