Electromyographic Responses Across Different Pulse-Widths of Sacral Neuromodulation in Sheep
Xin Su1, Matthew Cutinella2, Stephanie Koppes2
1Medtronic, Inc., Restorative Therapies Group, Research & Core Technology, Minneapolis, MN, USA.
Objectives:
In rodents, we reported that short pulse-width (PW) neuromodulation might provide more efficient therapy delivery than traditional 0.21 msec PW. Using fully implanted, commercialized systems in the sheep, the goal of this study was to characterize the relationship of electromyographic (EMG) responses of the external anal sphincter (EAS) to different PWs of sacral neuromodulation (SNM).
Materials And Methods:
In seven sheep, InterStim® quadripolar tined leads were implanted adjacent to the S3 nerve root bilaterally to deliver SNM and two pairs of intramuscular leads were placed on either side of the EAS for EMG sensing. The EMG responses to SNM with different PWs were examined using variable intensities in both anesthetized and conscious conditions.
Results:
The EMG responses from ipsilateral EAS (IEAS ) and contralateral EAS (CEAS ) were compared. The area under the curve of EMG responses from IEAS were significantly stronger than that from CEAS. The late component EMGs were more sensitive to nerve stimulation with a higher response amplitude in awake sheep. The response threshold-PW relationship from the IEAS as ascertained visually and with EMG in anesthetized and awake sheep were fitted with a monoexponential nonlinear regression; the resulting chronaxies were of 0.05 msec (n = 6), and 0.04 msec (n = 6), and 0.04 msec (n = 8), respectively.
Conclusions:
In both anesthetized and awake conditions, a similar motor response may be evoked in the EAS at PWs much shorter (0.04-0.05 msec) than the 0.21 msec typically used with SNM. Potential battery savings manifested by shorter PW would provide more efficient therapy delivery and increased longevity of the stimulator.
More Related Videos
Related Concept Videos
Cloning of Dolly the Sheep
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Regulation of Pulse