Volumetric Measurement of Subretinal Blebs Using Microscope-Integrated Optical Coherence Tomography
S Tammy Hsu1, Hesham Gabr1,2, Christian Viehland3
1Department of Ophthalmology, Duke University School of Medicine, Durham, NC, USA.
Purpose:
We advance studies of subretinal treatments by developing a microscope-integrated optical coherence tomography (MIOCT) image-based method for measuring the volume of therapeutics delivered into the subretinal space.
Methods:
A MIOCT image-based volume measurement method was developed and assessed for accuracy and reproducibility by imaging an object of known size in model eyes. This method then was applied to subretinal blebs created by injection of diluted triamcinolone. Bleb volumes obtained from MIOCT were compared to the intended injection volume and the surgeon's estimation of leakage.
Results:
Validation of the image-based volume measurement method showed accuracy to ±1.0 μL (6.0% of measured volume) with no statistically significant variation under different imaging settings. When this method was applied to subretinal blebs, four of 11 blebs without surgeon-observed leakage yielded a mean volume of 32 ± 12.5 μL, in contrast to the intended 50 μL volume injected from the delivery device. This constituted a mean difference of -18 μL (mean percent error, 36 ± 25%). For all 11 blebs, the surgeon's estimations of leakage were significantly different from and showed no correlation with the volume loss based on image-based volume measurements (P < 0.001, paired t-test; intraclass correlation = 0).
Conclusions:
We validated an accurate and reproducible method for measuring subretinal volumes using MIOCT. Use of this method revealed that the intended volume might not be delivered into the subretinal space. MIOCT can allow for accurate assessment of subretinal dose delivered, which may have therapeutic implications in evaluating the efficacy and toxicity of subretinal therapies.
Translational Relevance:
Use of MIOCT can provide feedback on the accuracy of subretinal injection volumes delivered.
Related Concept Videos
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Cell Motility through Blebbing
Blebbing Through the Matrix
In multicellular...
Integration by Parts: Indefinite Integrals
Integration by Parts: Definite Integrals
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Definite Integral