An experimental study of pulse wave measurements with magnetic induction phase shift method
Jian Sun1,2,1, Wei Zhuang2,1, Gen Li2,3
1Department of Neurosurgery, Southwest Hospital, Chongqing 400038, China.
Summary
Magnetic induction phase shift (MIPS) offers a more stable and accurate method for pulse wave monitoring compared to traditional techniques. This advanced sensor technology provides reliable physiological data for cardiovascular assessment.
Area of Science:
- Biomedical Engineering
- Cardiovascular Physiology
- Sensor Technology
Background:
- Pulse wave monitoring is crucial for assessing cardiovascular health.
- Existing methods require validation for improved accuracy and stability.
Purpose of the Study:
- To design high-sensitivity ring sensors for pulse detection.
- To develop a simultaneous acquisition platform using National Instruments T-Clock technology (NI-TCLK).
- To compare magnetic induction phase shift (MIPS) with traditional pulse detection methods.
Main Methods:
- Utilized a 10.7 MHz excitation signal with 20 dBm power.
- Employed a simultaneous acquisition platform with NI-TCLK technology.
- Recruited 30 adult volunteers (20-30 years) for sensor testing on forefingers.
Main Results:
- MIPS signal phase shift ranged from 0.6-0.8 degrees.
- MIPS demonstrated superior pulse rate stability and fewer deviations than the traditional method.
- Spectral analysis confirmed pulse frequencies between 1-3 Hz, with regularity around 1.5 Hz.
Conclusions:
- Magnetic induction pulse wave measurement accurately captures pulse wave characteristics.
- MIPS technology presents a viable advancement for cardiovascular monitoring.
- The developed system shows promise for clinical applications in pulse wave analysis.
Related Concept Videos
Inductance: Single-Phase And Three-Phase Line
643
Understanding the inductance of transmission lines is crucial for efficient design and operation in electrical power systems. This discussion delves into the inductance characteristics of single-phase two-wire and three-phase three-wire transmission lines with equal phase spacing.
Single-Phase Two-Wire Line:
A single-phase line consists of two solid cylindrical conductors, denoted as x and y. Each conductor carries phasor currents ix and iy, respectively. Given that the sum of these currents is...
Single-Phase Two-Wire Line:
A single-phase line consists of two solid cylindrical conductors, denoted as x and y. Each conductor carries phasor currents ix and iy, respectively. Given that the sum of these currents is...
643
Inductive Effects on Chemical Shift: Overview
2.2K
The protons in unsubstituted alkanes are strongly shielded with chemical shifts below 1.8 ppm. Methine, methylene, and methyl protons appear at approximately 1.7, 1.2 and 0.7 ppm, while the proton signal from methane appears at 0.23 ppm. An electronegative substituent, such as chlorine, withdraws the electron density from the protons, increasing their chemical shift. Progressive substitution of the hydrogens in methane by chlorine shifts the proton signals increasingly downfield, to 3.05 ppm in...
2.2K
Special considerations while measuring pulse
984
Assessing a patient's pulse is a fundamental skill in healthcare, but certain situations require special attention:
984
The Wave Nature of Light
61.6K
The nature of light has been a subject of inquiry since antiquity. In the seventeenth century, Isaac Newton performed experiments with lenses and prisms and was able to demonstrate that white light consists of the individual colors of the rainbow combined together. Newton explained his optics findings in terms of a "corpuscular" view of light, in which light was composed of streams of extremely tiny particles traveling at high speeds according to Newton's laws of motion.
61.6K
Pulse
2.2K
When the heart pumps blood out, arterial elastic fibers play a crucial role in sustaining a high-pressure gradient. They expand to accommodate the received blood and then recoil - a process known as the pulse that can be either manually palpated or electronically quantified. Despite a reduction in its effect with increased distance from the heart, elements of the pulse's systolic and diastolic components persist, observable even at the arteriole level.
The pulse serves as a clinical...
The pulse serves as a clinical...
2.2K
Pulse
4.2K
The pulse is one of the most fundamental physiological indicators of the body's cardiovascular health. It is the rhythmic expansion and contraction of the arterial walls in response to the pressure generated by the heart's pumping action.
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
4.2K