Related Experiment Video
Updated: Feb 11, 2026

Orthotopic Transplantation of Breast Tumors as Preclinical Models for Breast Cancer
Published on: May 18, 2020
Homogeneous vs. patient specific breast models for Monte Carlo evaluation of mean glandular dose in mammography
A Sarno1, G Mettivier1, F Di Lillo1
1Dipartimento di Fisica "Ettore Pancini", Università di Napoli Federico II & INFN Sez. di Napoli, I-80126 Napoli, Italy.
Purpose:
To compare, via Monte Carlo simulations, homogeneous and non-homogenous breast models adopted for mean glandular dose (MGD) estimates in mammography vs. patient specific digital breast phantoms.
Methods:
We developed a GEANT4 Monte Carlo code simulating four homogenous cylindrical breast models featured as follows: (1) semi-cylindrical section enveloped in a 5-mm adipose layer; (2) semi-elliptical section with a 4-mm thick skin; (3) semi-cylindrical section with a 1.45-mm skin layer; (4) semi-cylindrical section in a 1.45-mm skin layer and 2-mm subcutaneous adipose layer. Twenty patient specific digital breast phantoms produced from a dedicated CT scanner were assumed as reference in the comparison. We simulated two spectra produced from two anode/filter combinations. An additional digital breast phantom was produced via BreastSimulator software.
Results:
With reference to the results for patient-specific breast phantoms and for W/Al spectra, models #1 and #3 showed higher MGD values by about 1% (ranges [-33%; +28%] and [-31%; +30%], respectively), while for model #4 it was 2% lower (range [-34%; +26%]) and for model #2 -11% (range [-39%; +14%]), on average. On the other hand, for W/Rh spectra, models #1 and #4 showed lower MGD values by 2% and 1%, while for model #2 and #3 it was 14% and 8% lower, respectively (ranges [-43%; +13%] and [-41%; +21%]). The simulation with the digital breast phantom produced with BreastSimulator showed a MGD overestimation of +33%.
Conclusions:
The homogeneous breast models led to maximum MGD underestimation and overestimation of 43% and 28%, respectively, when compared to patient specific breast phantoms derived from clinical CT scans.
Related Concept Videos
Drug Dosing: Geriatric Patients
Drug Dosing: Obese Patients
Dose-Response Relationship: Selectivity and Specificity
Test for Homogeneity
Classification of Epithelial Tissues: Glandular Epithelium
Multicellular glands are formed during early development when epithelial budding...
Drug Excretion: Pulmonary and Glandular Routes
Drugs can also be excreted in breast milk, which is crucial for breastfeeding infants. The...