Introduction:
Insulation failure leading to conductor externalization (CE) of a Linox (Biotronik, Berlin, Germany) implantable cardioverter defibrillator (ICD) lead has recently been reported. The aim of this study was to assess prospectively all Linox family ICD leads implanted at our center for evidence of CE or an electrical abnormality.
Methods:
All patients with a Linox family ICD lead implanted at our center, between November 2007 and March 2015, were identified and all living patients were invited to attend for fluoroscopic screening and electrical assessment of the lead.
Results:
A total of 183 patients had a Linox family ICD lead implanted at our center. Of these, 5 patients (2.7%) had the lead extracted because of electrical failure and 2 of these leads had evidence of CE. Out of 158 living patients with a Linox family ICD lead, 111 patients attended for screening (mean age 63.1 years, 22.5% female). In this group of patients, no cases of CE or electrical abnormalities of the lead were identified.
Conclusion:
In this study evaluating 183 patients with a Linox family ICD lead implanted at a single center, 5 leads (2.7%) were explanted because of electrical failure and 2 of these leads had evidence of CE. Prospective fluoroscopic assessment of 111 Linox family ICD leads, with a mean dwell time of 31.5 months, revealed no further cases of CE.
More Related Videos
09:06Cochlear Implant Surgery and Electrically-evoked Auditory Brainstem Response Recordings in C57BL/6 Mice
Published on: January 9, 2019
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Related Concept Videos
Abnormal Proliferation
Cardiovascular System Abnormal Findings I: Inspection and Palpation
Abnormal findings observed during an inspection
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Phase-lead and Phase-lag Controllers
Structural Protein Function
Collagen, the most abundant protein in mammals, is found throughout the body. In connective tissue, such as skin, ligaments, and tendons, it provides tensile strength and elasticity. In bones and teeth, it mineralizes to...
Cardiovascular System Abnormal Findings II: Auscultation
Abnormal Heart Sounds
Gallops: