Imaging Metals in Brain Tissue by Laser Ablation - Inductively Coupled Plasma - Mass Spectrometry LA-ICP-MS
Published on: January 22, 2017
Quantitative Bioimaging of Platinum via Online Isotope Dilution-Laser Ablation-Inductively Coupled Plasma Mass
Oliver Bolle Bauer1, Christina Köppen1, Michael Sperling1,2
1Westfälische Wilhelms-Universität Münster , Institute of Inorganic and Analytical Chemistry , Corrensstraße 30 , 48149 Münster , Germany.
Abstract:
A new calibration strategy for elemental bioimaging based on online isotope dilution analysis (IDA) and laser ablation-inductively coupled plasma mass spectrometry (LA-ICPMS) was developed and applied for the quantification of platinum in rat kidney tissues. A dry 194Pt spike aerosol was added in a post-cell setup, and the natural 194Pt/195Pt isotope ratio of the sample aerosol from laser ablation was changed accordingly. Spike mass flow determination was carried out based on reversed IDA using a reference standard. Quantitative data obtained by the new approach correlated well with those obtained by external calibration when analyzing parallel tissue slices of rat kidney from cisplatin perfusion studies. The novel quantification approach is traceable to SI units, as IDA is an definitive method. Signal drifts are compensated as the second isotope acts as an internal standard.
More Related Videos
07:25Quantitative Metabolomics of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: January 5, 2021
08:46Dose Uptake of Platinum- and Ruthenium-based Compound Exposure in Zebrafish by Inductively Coupled Plasma Mass Spectrometry with Broader Applications
Published on: April 21, 2022
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometry: Isotope Effect
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Isotopes
An element's atomic mass, or weight,...