Improving the sensitivity of miniature linear ion trap mass spectrometer by a DC voltage applied on the eject
Gang Li1, Detian Li1, Yongjun Cheng1
1Science and Technology on Vacuum Technology and Physics Laboratory, Lanzhou Institute of Physics, Lanzhou, China.
Abstract:
A miniaturized linear ion trap mass spectrometer with continuous atmospheric pressure interface has been built in our lab. Significant extension in mass range and reduction in power consumption have been realized by the supplemental alternating current frequency scan mode. However, relatively poor sensitivity has been witnessed, which is directly dominated by the detection efficiency of the ion detector. Theoretical analysis has been implemented to find ways to improve the detection efficiency. The results show that enhanced sensitivity can be obtained by applying a direct current voltage on the pair of electrodes in eject direction. Experiments show that the sensitivity has been improved by more than one time due to the application of direct current voltage. With this design, this homemade miniature linear ion trap mass spectrometer can be used to analyze more rarefied samples, especially to on-site chemical analysis and space application.
Related Concept Videos
Mass Spectrometers
Voltage-gated Ion Channels
Generally, all voltage-gated ion channels have a 'voltage-sensing domain' that spans the lipid bilayer. The charged residues in the sensor move in response to the membrane potential changes that open the channel allowing ions movement. There are several types of...
IR Spectrometers
Standard Electrode Potentials
DC Generator
Ion Channels
Ion channels are specialized integral membrane proteins on the plasma membrane that allow...