Cusp Tear of Trifecta Aortic Bioprosthesis Resulting in Acute Heart Failure
Anne-Kristin Schaefer1, Alfred Kocher1, Günther Laufer1
1Division of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.
Abstract:
Herein is presented the case of an 83-year-old male patient in cardiogenic shock with acute aortic regurgitation that occurred six years after aortic valve replacement (AVR) with a 23 mm Trifecta™ valve. Prosthesis endocarditis was initially suspected because of a floating structure attached to the aortic valve that was visible on echocardiography. Emergency redo-AVR surgery was performed, but no signs of endocarditis were found intraoperatively. Hence, cusp tearing of the implanted bioprosthesis was considered to be the reason for the severe aortic regurgitation.
More Related Videos
04:05Author Spotlight: Unveiling Prognostic Indicators in Heart Failure - The Role of Phase Angle and Bioelectrical Impedance Analysis
Published on: June 30, 2023
07:49Author Spotlight: Investigating HR-Dependent Cardiac Function in Mouse Models Through a Novel Atrial-Pacing Approach
Published on: July 21, 2023
Related Concept Videos
Acute Respiratory Failure-I
Definition: It is defined by specific criteria based on blood gas measurements. Hypoxemia happens when the partial pressure of oxygen (PaO2) falls below 60 mmHg. At the same time,...
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Heart Failure II: Pathophysiology
Pathophysiology of Heart Failure
Heart Failure I: Introduction