Consistency of Continuous Ambulatory Interstitial Glucose Monitoring Sensors
Pei T Wu1, David E Segovia2, Cathy C Lee3
1Division of Cardiology, David Geffen School of Medicine at UCLA and VA Greater Los Angeles Healthcare System, Los Angeles, CA 90024, USA. ptwu@ucla.edu.
Aims:
The abdominal region is the most common location for continuous glucose monitor (CGM) sensor insertion. However, a paucity of post-marketing data is available to demonstrate intra-individual consistency of CGM readings at different abdominal insertion sites.
Methods:
Healthy adults (fasting glucose (FG) < 5.5 mmol/L; BMI < 30 kg/m²) were recruited and a CGM sensor was placed on each side of the abdomen. Postprandial and continuous 48-h interstitial glucose levels were analyzed.
Results:
There was no significant difference in the 3-h postprandial glucose (PPG) level derived from the left versus right CGM, which remained non-significant after adjusting for waist circumference or FG. Among the glucose levels recorded over 48-h, values on the left site were greater in 3.6% of the data points (p < 0.05). After adjusting for waist circumference, only 0.5% of the glucose values remained significantly greater on the left (p < 0.05). When adjusted for FG, similar results were observed. For both PPG and 48-h readings, the mean absolute relative difference was not significant between the two abdominal sites.
Conclusions:
CGM-derived glucose measures were highly consistent between the left and right abdomen during both the postprandial and post-absorptive periods.
Related Concept Videos
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Glucose Transporters
Facilitated diffusion-glucose transporters (GLUTs) are encoded by the solute-linked carrier (SLC) family 2, subfamily A gene family, or SLC2A. The 14 GLUT protein members are distributed into three classes:
Continuing Care
Continuity of a Function
Continuity Equation
The mass flow rate is expressed as:
Continuity Equation