Halving the Volume of AnaConDa: Evaluation of a New Small-Volume Anesthetic Reflector in a Test Lung Model
Hagen Bomberg1, Franziska Meiser1, Philipp Daume1
1From the Department of Anesthesiology, Intensive Care Medicine and Pain Medicine, Saarland University Medical Centre, University of Saarland, Homburg/Saar, Germany.
Background:
Volatile anesthetics are increasingly used for sedation in intensive care units. The most common administration system is AnaConDa-100 mL (ACD-100; Sedana Medical, Uppsala, Sweden), which reflects volatile anesthetics in open ventilation circuits. AnaConDa-50 mL (ACD-50) is a new device with half the volumetric dead space. Carbon dioxide (CO2) can be retained with both devices. We therefore compared the CO2 elimination and isoflurane reflection efficiency of both devices.
Methods:
A test lung constantly insufflated with CO2 was ventilated with a tidal volume of 500 mL at 10 breaths/min. End-tidal CO2 (EtCO2) partial pressure was measured using 3 different devices: a heat-and-moisture exchanger (HME, 35 mL), ACD-100, and ACD-50 under 4 different experimental conditions: ambient temperature pressure (ATP), body temperature pressure saturated (BTPS) conditions, BTPS with 0.4 Vol% isoflurane (ISO-0.4), and BTPS with 1.2 Vol% isoflurane. Fifty breaths were recorded at 3 time points (n = 150) for each device and each condition. To determine device dead space, we adjusted the tidal volume to maintain normocapnia (n = 3), for each device. Thereafter, we determined reflection efficiency by measuring isoflurane concentrations at infusion rates varying from 0.5 to 20 mL/h (n = 3), for each device.
Results:
EtCO2 was consistently greater with ACD-100 than with ACD-50 and HME (ISO-0.4, mean ± standard deviations: ACD-100, 52.4 ± 0.8; ACD-50, 44.4 ± 0.8; HME, 40.1 ± 0.4 mm Hg; differences of means of EtCO2 [respective 95% confidence intervals]: ACD-100 - ACD-50, 8.0 [7.9-8.1] mm Hg, P < .001; ACD-100 - HME, 12.3 [12.2-12.4] mm Hg, P < .001; ACD-50 - HME, 4.3 [4.2-4.3] mm Hg, P < .001). It was greatest under ATP, less under BTPS, and least with ISO-0.4 and BTPS with 1.2 Vol% isoflurane. In addition to the 100 or 50 mL "volumetric dead space" of each AnaConDa, "reflective dead space" was 40 mL with ACD-100 and 25 mL with ACD-50 when using isoflurane. Isoflurane reflection was highest under ATP. Under BTPS with CO2 insufflation and isoflurane concentrations around 0.4 Vol%, reflection efficiency was 93% with ACD-100 and 80% with ACD-50.
Conclusions:
Isoflurane reflection remained sufficient with the ACD-50 at clinical anesthetic concentrations, while CO2 elimination was improved. The ACD-50 should be practical for tidal volumes as low as 200 mL, allowing lung-protective ventilation even in small patients.
Related Concept Videos
Constant Volume Calorimetry
Volume of Distribution
Problem Solving: Volume
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Work Done During Volume Change
Consider a gas confined to a cylinder fitted with a movable piston at one end. If the gas expands from volume V1 to volume V2, it exerts a force on the piston, such that the piston moves by a distance dr.
The work done by the gas on the piston can be expressed as
Control Volume and System Representations
The control volume approach considers a stationary region in space through which fluid flows. This region is bounded by a control surface. For instance, in the case of water...