MASE-sLASER, a short-TE, matched chemical shift displacement error sequence for single-voxel spectroscopy at
Seyedmorteza Rohani Rankouhi1,2, Donghyun Hong1, Hadrien Dyvorne3
1Erwin L. Hahn Institute for Magnetic Resonance Imaging, University of Duisburg-Essen, Essen, Germany.
Abstract:
B1 inhomogeneity and chemical shift displacement error (CSDE) increase with the main magnetic field strength and are therefore deleterious for magnetic resonance spectroscopy (MRS) at ultrahigh field. A solution is to use adiabatic pulses which operate over a broad range of B1 and thus are insensitive to B1 inhomogeneity. Moreover, adiabatic pulses usually have a relatively higher bandwidth, which makes CSDE low to negligible. The use of exclusively adiabatic pulses for single-voxel spectroscopy (SVS) typically brings the disadvantage of a long echo time (TE), but the advantage of a low and matched CSDE. Herein, we took advantage of short-duration, low-power, matched-phase adiabatic spin echo (MASE) pulses to implement a matched CSDE semi-localized by adiabatic selective refocusing (sLASER) sequence capable of attaining short TEs, while CSDE is matched and still comparatively low. We also demonstrate here the feasibility of the direct measurement of the γ-aminobutyric acid (GABA) resonance at 2.28 ppm well separated from the neighboring glutamate resonance at 7 T using the implemented MASE-sLASER sequence at TEs of 68 and 136 ms. The shorter duration of MASE pulses also made it possible to implement a Mescher-Garwood-semi-localized by adiabatic selective refocusing (MEGA-sLASER) (with MASE) sequence with TE = 68 ms for editing GABA at 7 T, the results for which are also shown.
Related Concept Videos
NMR Spectroscopy: Chemical Shift Overview
For instance, the proton...
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
Inductive Effects on Chemical Shift: Overview
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
π Electron Effects on Chemical Shift: Overview
¹H NMR Chemical Shift Equivalence: Homotopic and Heterotopic Protons