Totally implanted device for venous access. Experience in tumour patients
Acta Radiologica. Oncology
|March 1, 1985
Summary
The Port-A-Cath system provides reliable central venous access for chemotherapy in cancer patients. This implantable device demonstrated excellent functionality and patient acceptance with minimal complications over an 18-month study period.
Area of Science:
- Oncology
- Medical Devices
- Vascular Access
Background:
- Central venous access is crucial for administering chemotherapy and supportive care in cancer patients.
- Implantable devices like the Port-A-Cath offer a long-term solution for repeated venous access.
- Managing complications and ensuring device patency are key considerations for such systems.
Purpose of the Study:
- To evaluate the safety, efficacy, and patient acceptance of the Port-A-Cath system for central venous access in tumor patients.
- To assess the device's functionality and complication rates during an 18-month investigation period.
Main Methods:
- Implantation of a subcutaneous injection portal (Port-A-Cath) connected to a silicone catheter in 30 cancer patients.
- Utilized the system primarily for administration of antineoplastic drugs and continuous infusions.
- Monitored device function, drug administration frequency, infusion durations, and complications over 18 months.
Main Results:
- The Port-A-Cath system remained functional for a median of 155 days (range 7-372 days) in 30 patients.
- Drug injections administered to 22 patients (median 11 occasions), and continuous infusions to 17 patients (median 15 days).
- Low complication rates: one catheter extravasation, two skin erosions, and one case of caval thrombosis; no infections or occlusions reported. Ten patients died with functioning systems, and 18 had working devices at study completion.
Conclusions:
- The Port-A-Cath system is a safe and effective method for long-term central venous access in cancer patients undergoing chemotherapy.
- The device exhibits high functionality and excellent patient acceptance with a low incidence of serious complications.
- Routine flushing and careful monitoring contribute to maintaining device patency and minimizing adverse events.