Related Experiment Video
Updated: Feb 9, 2026

Reducing Willow Wood Fuel Emission by Low Temperature Microwave Assisted Hydrothermal Carbonization
Published on: May 19, 2019
[Improvement and Application of Radiated Emission Test Method for Medical Microwave Equipment]
Zhanshuo Wang1, Wei Xu1, Zhiwei Hu1
1Tianjin Medical Devices Quality Supervision and Testing Center, Tianjin, 300384. ##Email#.
Abstract:
Distortion will occur when testing medical microwave products, due to its high power, high frequency, in the existing radiated emission testing process which will cause saturation non-linear of the pre-amplifier. It will lead to test distorted, the test results are not accurate. This article will improve the original test method by adding some suppressing-frequency devices to suppress the main frequency energy and avoid pre-amplifier saturation, eliminate distortion. All of this will ensure the test results are effective and relia ble.