Highly crosslinked polyethylene wear rates and acetabular component orientation: a minimum ten-year follow-up
M G Teeter1, B A Lanting1, D D Naudie1
1Department of Surgery, Schulich School of Medicine and Dentistry, Western University and London Health Sciences Centre, London, Ontario, Canada.
Aims:
The aim of this study was to determine whether there is a difference in the rate of wear between acetabular components positioned within and outside the 'safe zones' of anteversion and inclination angle.
Patients And Methods:
We reviewed 100 hips in 94 patients who had undergone primary total hip arthroplasty (THA) at least ten years previously. Patients all had the same type of acetabular component with a bearing couple which consisted of a 28 mm cobalt-chromium head on a highly crosslinked polyethylene (HXLPE) liner. A supine radiostereometric analysis (RSA) examination was carried out which acquired anteroposterior (AP) and lateral paired images. Acetabular component anteversion and inclination angles were measured as well as total femoral head penetration, which was divided by the length of implantation to determine the rate of polyethylene wear.
Results:
The mean anteversion angle was 19.4° (-15.2° to 48°, sd 11.4°), the mean inclination angle 43.4° (27.3° to 60.5°, sd 6.6°), and the mean wear rate 0.055 mm/year (sd 0.060). Exactly half of the hips were positioned inside the 'safe zone'. There was no difference (median difference, 0.012 mm/year; p = 0.091) in the rate of wear between acetabular components located within or outside the 'safe zone'. When compared to acetabular components located inside the 'safe zone', the wear rate was no different for acetabular components that only achieved the target anteversion angle (median difference, 0.012 mm/year; p = 0.138), target inclination angle (median difference, 0.013 mm/year; p = 0.354), or neither target (median difference, 0.012 mm/year; p = 0.322).
Conclusion:
Placing the acetabular component within or outside the 'safe zone' did not alter the wear rate of HXLPE at long-term follow-up to a level that risked osteolysis. HXLPE appears to be a forgiving bearing material in terms of articular surface wear, but care must still be taken to position the acetabular component correctly so that the implant is stable. Cite this article: Bone Joint J 2018;100-B:891-7.
Related Concept Videos
Temperature Dependence on Reaction Rate
Atoms, molecules, or ions must collide before they can react with each other. Atoms must be close together to form chemical bonds. This premise is the basis for a theory that explains many observations regarding chemical kinetics, including factors affecting reaction rates.
The collision theory is based on the postulates that (i) the reaction rate is proportional to the rate of reactant collisions, (ii) the reacting species collide in an orientation allowing contact between...
Location and Orientation of the Heart
Measuring Reaction Rates
Components of Stress
Interestingly, the hidden cube faces also experience these stresses, equal and...
Components of Language
Reaction Rate
The mathematical representation of the change in the concentration of reactants and products, over time, is the rate...