Iterative algorithms for metal artifact reduction in children with orthopedic prostheses: preliminary results
Seema Toso1, Meryle Laurent2, Elise Dupuis Lozeron3
1Paediatric Radiology Unit, Division of Radiology, Geneva Children's Hospitals, Rue Willy-Donzé 6, 1211, Geneva, Switzerland. seema.toso@hcuge.ch.
Background:
Increased computational power allows computed tomography (CT) software to process very advanced mathematical algorithms to generate better quality images at lower doses. One such algorithm, iterative metal artifact reduction (iMAR) has proven to decrease metal artifacts seen in CT images of adults with orthopedic implants.
Objectives:
To evaluate artifact reduction capability of the algorithm in lower-dose pediatric CT compared to our routine third-generation advanced modeled iterative reconstruction (ADMIRE) algorithm.
Materials And Methods:
Thirteen children (11-17 years old) with metal implants underwent routine clinically indicated CT. Data sets were reconstructed with an iMAR algorithm. Hounsfield units and image noise were measured in bone, muscle and fat in the streak artifact (near the implant) and at the greatest distance from the artifact (far from the implant). A regression model compared the effects of the algorithm (standard ADMIRE vs. iMAR) near and far from the implant.
Results:
Near the implant, Hounsfield units with iMAR were significantly different in our standard ADMIRE vs. iMAR for bone, muscle and fat (P<0.001). Noise was significantly different in standard ADMIRE vs. iMAR in bone (P<0.003). Far from the implant, Hounsfield units and noise were not significantly different for ADMIRE vs. iMAR, for the three tissue types.
Conclusion:
These preliminary results demonstrate that iMAR algorithms improves Hounsfield units near the implant and decreases image noise in bone in low-dose pediatric CT. It does this without changing baseline tissue density or noise far from the implant.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
05:42Orthopedic Robot-Assisted Femoral Neck System in the Treatment of Femoral Neck Fracture
Published on: March 3, 2023
Related Concept Videos
Oxidation-Reduction Reactions
Metal-Ligand Bonds
In these complexes, transition metals form coordinate covalent bonds, a kind of Lewis acid-base interaction in which both of the electrons in the bond are contributed by a donor (Lewis base) to an electron acceptor (Lewis acid). The Lewis acid in...
Bonding in Metals
Metallic Solids
All metallic solids exhibit high thermal and electrical conductivity, metallic luster, and malleability....
Alkali Metals
Table 1: Properties of the alkali metals
Trial and Error and Algorithm