Diagnostic accuracy of third-generation dual-source dual-energy CT: a prospective trial and protocol for clinical
Tim Nestler1, Kai Nestler2, Andreas Neisius3
1Department of Urology, Federal Armed Services Hospital Koblenz, Rübenacher Str. 170, 56072, Koblenz, Germany. tim-nestler@web.de.
Purpose:
Uric acid (UA) calculi can be referred to chemolitholysis rather than invasive treatment. Dual-energy computed tomography (DECT) may be able to distinguish between UA and non-UA (NUA) calculi. The aim of this study was to evaluate the validity of third-generation DECT for the first time and to investigate whether combining DECT with clinical parameters can increase its predictive accuracy.
Materials And Methods:
All patients who presented to our emergency department between January 2015 and March 2017 with urinary stones were prospectively included in this observational study and underwent DECT with subsequent interventional stone removal. Stone composition was analyzed using infrared spectrometry as the gold standard. Predictive accuracy of DECT and clinical covariates was computed by assessing univariate and multivariate areas under the curve (AUCs).
Results:
Of 84 patients with 144 urinary stones, 10 (11.9%) patients had UA stones according to infrared spectrometry, and the remaining stones were NUA or mixed stones. DECT had a positive predictive value of 100% and a negative predictive value of 98.5% for UA stones. The AUC for urine pH alone was 0.71 and 0.97 for DECT plus urine pH. No UA stones were found in patients with a urine pH above > 5.5. Mean DLP was 225.15 ± 128.60 mGy*cm and mean effective dose was 3.38 ± 1.93 mSv.
Conclusions:
DECT is a safe method for assigning patients to oral chemolitholysis. Clinical preselection of patients based on urinary pH (< 6.0) leads to a more liable use of DECT. Third-generation DECT needs significant lower radiation doses compared to previous generations.
Related Concept Videos
Clinical Trials
There are four phases in a clinical trial. A phase one...
Clinical Trials: Overview
Dual Nature of Electromagnetic (EM) Radiation
Wavelength is the distance between two consecutive peaks (the highest point) or troughs (the lowest point) in the wave. Frequency is the number of...
Pericarditis II: Clinical Features and Diagnostic Tests
Atherosclerosis II: Clinical Manifestations and Diagnostic Tests
Aneurysm II: Clinical Manifestations and Diagnostic Studies