Effect of CT Acquisition Parameters on Iodine Density Measurement at Dual-Layer Spectral CT
Hyungjin Kim1, Chang Min Park1,2, Chang Kyung Kang3
11 Department of Radiology, Seoul National University College of Medicine, and Institute of Radiation Medicine, Seoul National University Medical Research Center, 101 Daehak-ro, Jongno-gu, Seoul, 03080, Korea.
Objective:
We aimed to evaluate the effect of tube voltage, tube current-time product, and iterative reconstruction on iodine quantification using a dual-layer spectral CT scanner.
Materials And Methods:
Two mediastinal iodine phantoms, each containing six tubes of different iodine concentrations (0, 1, 2.5, 5, 10, and 20 mg I/mL; the two phantoms had tubes with contrast media diluted in water and in 10% amino acid solution, respectively), were inserted into an anthropomorphic chest phantom and scanned with varying acquisition parameters (120 and 140 kVp; 20, 40, 60, 80, 100, 150, and 200 mAs; and spectral reconstruction levels 0 and 6). Thereafter, iodine density was measured (in milligrams of iodine per milliliter) using a dedicated software program, and the effect of acquisition parameters on iodine density and on its relative measurement error (RME) was analyzed using a linear mixed-effects model.
Results:
Tube voltages (all, p < 0.001) and tube current-time products (p < 0.05, depending on the interaction terms for iodine density; p = 0.023 for RME) had statistically significant effects on iodine density and RME. However, the magnitude of their effects was minimal. That is, estimated differences between tube voltage settings ranged from 0 to 0.8 mg I/mL for iodine density and from 1.0% to 4.2% for RME. For tube current-time product, alteration of 100 mAs caused changes in iodine density and RME of approximately 0.1 mg I/mL and 0.6%, respectively. Spectral level was not an affecting factor for iodine quantification (p = 0.647 for iodine density and 0.813 for RME).
Conclusion:
Iodine quantification using dual-layer spectral CT was feasible irrespective of CT acquisition parameters because their effects on iodine density and RME were minimal.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Wave Parameters
Electrophilic Aromatic Substitution: Fluorination and Iodination of Benzene