Inappropriate ICD shock due to hot tub-induced external electrical interference
Mohammad Shenasa1, Shahriar Heidary1, Hossein Shenasa1
1Heart & Rhythm Medical Group, San Jose, CA 95128, USA; O'Connor Hospital, San Jose, CA 95128, USA.
Background:
A 72-year-old white male with a history of rapid nonsustained ventricular tachycardia, hypertrophic cardiomyopathy, and intermittent Brugada-type ECG had a single-lead implantable cardioverter-defibrillator (ICD) implantation and received a sudden ICD shock while in the hot tub. To the best of our knowledge this is the first case report of hot tub jet-induced inappropriate ICD shock.
Methods:
ICD interrogation and analysis of intracardiac electrograms and event markers.
Results:
ICD interrogation revealed inappropriate ICD shocks due to electrical interference of hot tub engine; 60-cycle electrical artifact mimicking fast ventricular fibrillation erroneously detected by the device. The device then delivered a 34.8 joules shock while the patient was actually in sinus rhythm.
Conclusions:
Electrical interference due to external sources such as hot tub engines may occur and produce an inappropriate detection and ICD shock. Precaution and patient education is warranted.
Related Concept Videos
RNA Interference
This process occurs naturally in cells, often through the activity of genomically-encoded microRNAs. Researchers can take advantage of this mechanism by introducing synthetic RNAs to deactivate specific genes for research or therapeutic purposes. For example, RNAi could be used...
Induced Electric Dipoles
Since the absolute value of potential energy holds no physical meaning, its zero value can be chosen as per...
Induced Electric Fields
Interference and Diffraction
Induced Electric Fields: Applications
Hot Weather Concreting
Mitigating the heat increase in concrete can be economically achieved by shading aggregate stockpiles to prevent heating from solar radiation,...