Intrauterine Pressure During Hysteroscopic Morcellation: A Comparison of Three Commercially-Available Devices
Erica Stockwell1, David L Howard2
1Minimally Invasive Gynecologic Surgeon, Las Vegas Minimally Invasive Surgery Well Health Quality Care, a DaVita Medical Group, Department of Obstetrics and Gynecology, University of Nevada, Las Vegas School of Medicine, Las Vegas, Nevada.
Study Objective:
Our objective was to compare intrauterine pressures during resection and aspiration modes among three types of commercially-available hysteroscopic morcellators.
Design:
This was a benchtop study (Canadian Task Force level II-1). This study cannot feasibly and ethically be done in-vivo, so an ex-vivo study design was chosen.
Setting:
A silicone uterine model was attached to a manometer via tubing, with the tip inside the cavity to allow for intracavity pressure measurements. Each hysteroscopic morcellator was then introduced, and intracavity pressures were recorded every one to two seconds in three modes (static, resection, and aspiration) and at three set point pressures (45, 85, and 125 mmHg).
Patients:
No human subjects were involved in this study.
Interventions:
None.
Measurements And Main Results:
There were a total of 4,872 pressure measurements during this study across the three devices, over the three modes, and at the three set point pressures combined. Using mixed-effects linear regression, the mean observed intracavity pressure was not greater than the set pressure for each of the three devices. This result held true in both aspiration and resection modes. In our statistical models, the coefficient on the terms representing the interaction between device and time were not statistically significant in either resection or aspiration modes. This indicates that, statistically, the change in intracavity pressure over time was not significantly different across the three devices.
Conclusion:
In this first of its kind head-to-head benchtop study, we found that all three commercially-available hysteroscopic morcellators appear to be similar to each other in terms of their abilities to maintain intracavity pressure below the set pressure, which is important in avoiding intravasation in-vivo. These findings are important because many gynecologists do not have the ability to choose between the three available devices on the market at their institution.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Vapor Pressure
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...