Design and construction of an innovative brain phantom prototype for MRI
Anna Altermatt1,2, Francesco Santini1,3, Xeni Deligianni1,3
1Medical Image Analysis Center (MIAC) AG, Basel, Switzerland.
Purpose:
The purpose of this project was to construct a physical brain phantom for MRI, mimicking structure and T1 relaxation properties of white matter (WM) and gray matter (GM).
Methods:
The phantom design comprised 2 compartments, 1 resembling the WM and 1 resembling the GM. Their T1 relaxation times, as assessed using an inversion recovery turbo spin echo sequence, were reproduced using an agar gel doped with contrast agent (CA) and their folding patterns were simulated through a molding-casting procedure using 3D-printed casts and flexible silicone molds. Three versions of the assembling procedure were adopted to build: Phantom1 without any separation; Phantom2 with a varnish layer; and Phantom3 with a thin wax layer between the compartments.
Results:
Phantom1 was characterized by an immediate diffusion of CA between the 2 compartments. Phantom2 and Phantom3, instead, showed relaxation times and shape comparable with the target ones identified in a healthy control subject (WM: 754 ± 40 ms; GM: 1277 ± 96 ms). Moreover, both compartments revealed intact gyri and sulci. However, the diffusion of CA made Phantom2 stable only for a short period of time. Phantom3 showed stability within a time window of several days but the wax layer between the WM and GM was visible in the MRI.
Conclusion:
Structural and intensity properties of the constructed phantoms are useful in evaluating and validating steps from image acquisition to image processing. Moreover, the described constructing procedure and its modular design make it adjustable to a variety of applications.
More Related Videos
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
11:29Real-time Video Projection in an MRI for Characterization of Neural Correlates Associated with Mirror Therapy for Phantom Limb Pain
Published on: April 20, 2019
Related Concept Videos
Design Example: Dimensioning of Concrete Masonry Construction
The site engineer has laid out a plan for the storeroom with external dimensions of twelve feet in length and...
Concepts and Prototypes
The brain organizes this information using concepts, which are mental categories grouping linguistic data,...
Group Design
Factorial Design
Design Example: Designing a Residential Plumbing System
Design Example: Designing Water Slide
Bernoulli's principle determines the water's velocity along the slide....