Related Experiment Video
Updated: Feb 4, 2026

Robotic Cochlear Implantation for Direct Cochlear Access
Published on: June 16, 2022
Safety of radiofrequency ablation for adenotonsillectomy after cochlear implantation
Matthew C Gropler1, Brittany A Leader2, David J Brown3
1Department of Otolaryngology - Head and Neck Surgery, University of Cincinnati School of Medicine, Cincinnati, OH, United States; Department of Otolaryngology - Head and Neck Surgery, Saint Louis University School of Medicine, St. Louis, MO, United States.
Objective:
While a cadaveric animal study has suggested that radiofrequency ablation can be safely used in patients with cochlear implants, no in vivo studies have been published to confirm that radiofrequency ablation does not alter the integrity of the cochlear implant device.
Methods:
Cochlear implant impedance and functional performance were studied through a prospective case series in five children with seven functioning multichannel implants before and after radiofrequency ablation adenotonsillectomy.
Results:
There were 4 females and 1 male patient, aged 6-10 years (mean 8.5 ± 1.95 years) with 7 functioning implants. Pre- and post-surgical impedance testing revealed all electrodes were within normal operating limits. There was no statistically significant difference between the mean pre and post-operative impedances in 5 of the 7 tested implants (P = 0.2-0.8). The other two implants showed statistically significant improvement in impedance values which were not clinically significant (P = 0.02 and P < 0.001). Speech perception was unchanged as was functional performance for all 7 tested implants.
Conclusions:
We found that radiofrequency ablation used in the oropharynx during adenotonsillectomy did not alter the integrity of the cochlear implant devices when assessed using electrode impedance testing, audiometry and speech perception evaluation. These results confirm those reported in previous in vitro studies and confirm the safety of radiofrequency ablation adenotonsillectomy for children who have undergone previous cochlear implant placement.
Related Concept Videos
Survey Safety
Household Wiring And Electrical Safety
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Cleavage and Blastulation
Reproductive Cloning
Somatic Cell Nuclear Transfer
In SCNT, an egg cell is taken from an animal and its nucleus is removed, creating an enucleated egg. Then a somatic...
Bacterial Signaling