A simple geometric analysis method for measuring and mitigating RF induced currents on Deep Brain Stimulation leads
Yigitcan Eryaman1, Naoharu Kobayashi1, Sean Moen1
1Center for Magnetic Resonance Research, University of Minnesota, Minneapolis, MN, United States.
Abstract:
The purpose of this work is to present a new method that can be used to estimate and mitigate RF induced currents on Deep Brain Stimulation (DBS) leads. Here, we demonstrate the effect of RF induced current mitigation on both RF heating and image quality for a variety of brain MRI sequences at 3 T. We acquired pre-scan images around a DBS lead (in-situ and ex-vivo) using conventional Gradient Echo Sequence (GRE) accelerated by parallel imaging (i.e GRAPPA) and quantified the magnitude and phase of RF induced current using the relative location of the B1+ null with respect to the lead position. We estimated the RF induced current on a DBS lead implanted in a gel phantom as well as in a cadaver head study for a variety of RF excitation patterns. We also measured the increase in tip temperature using fiber-optic probes for both phantom and cadaver studies. Using the magnitude and phase information of the current induced separately by two transmit channels of the body coil, we calculated an implant friendly (IF) excitation. Using the IF excitation, we acquired T1, T2 weighted Turbo Spin Echo (TSE), T2 weighted SPACE-Dark Fluid, and Ultra Short Echo Time (UTE) sequences around the lead. Our induced current estimation demonstrated linear relationship between the magnitude of the induced current and the square root SAR at the tip of the lead as measured in phantom studies. The "IF excitation pattern" calculated after the pre-scan mitigated RF artifacts and increased the image quality around the lead. In addition, it reduced the tip temperature significantly in both phantom and cadaver studies compared to a conventional quadrature excitation while keeping equivalent overall image quality. We present a relatively fast method that can be used to calculate implant friendly excitation, reducing image artifacts as well as the temperature around the DBS electrodes. When combined with a variety of MR sequences, the proposed method can improve the image quality and patient safety in clinical imaging scenarios.
More Related Videos
09:16A General Method for Evaluating Deep Brain Stimulation Effects on Intravenous Methamphetamine Self-Administration
Published on: January 22, 2016
09:46Analysis of Gene Expression Changes in the Rat Hippocampus After Deep Brain Stimulation of the Anterior Thalamic Nucleus
Published on: March 8, 2015
Related Concept Videos
Geometric Mean
In cases of multiplicative data, the geometric mean is used for statistical analysis. First, the product of all the elements is taken. Then, if there are n elements in the...
Geometric Sequences
Mesh Analysis with Current Sources
Current Source in One Mesh: The analysis process is straightforward when a current source is found in only one mesh within the circuit. Mesh currents are assigned as usual, with the mesh containing the current source excluded from the analysis. Kirchhoff's voltage law...
Design of Transmission Shafts - Stress Analysis
Performing a Simple Data Analysis using MS-Excel Function
SUM: This function calculates the total sum of a range of values. It's the foundation for aggregating data, essential for determining overall trends and totals in datasets.
AVERAGE: It computes the mean value of a given set of numbers, providing a quick insight into the central...
Phase-lead and Phase-lag Controllers