A regional bolus tracking and real-time B1 calibration method for hyperpolarized 13 C MRI
Shuyu Tang1, Eugene Milshteyn1, Galen Reed2
1UC Berkeley-UCSF Graduate Program in Bioengineering, University of California, San Francisco and University of California, Berkeley, California.
Purpose:
Acquisition timing and B1 calibration are two key factors that affect the quality and accuracy of hyperpolarized 13 C MRI. The goal of this project was to develop a new approach using regional bolus tracking to trigger Bloch-Siegert B1 mapping and real-time B1 calibration based on regional B1 measurements, followed by dynamic imaging of hyperpolarized 13 C metabolites in vivo.
Methods:
The proposed approach was implemented on a system which allows real-time data processing and real-time control on the sequence. Real-time center frequency calibration upon the bolus arrival was also added. The feasibility of applying the proposed framework for in vivo hyperpolarized 13 C imaging was tested on healthy rats, tumor-bearing mice and a healthy volunteer on a clinical 3T scanner following hyperpolarized [1-13 C]pyruvate injection. Multichannel receive coils were used in the human study.
Results:
Automatic acquisition timing based on either regional bolus peak or bolus arrival was achieved with the proposed framework. Reduced blurring artifacts in real-time reconstructed images were observed with real-time center frequency calibration. Real-time computed B1 scaling factors agreed with real-time acquired B1 maps. Flip angle correction using B1 maps results in a more consistent quantification of metabolic activity (i.e, pyruvate-to-lactate conversion, kPL ). Experiment recordings are provided to demonstrate the real-time actions during the experiment.
Conclusions:
The proposed method was successfully demonstrated on animals and a human volunteer, and is anticipated to improve the efficient use of the hyperpolarized signal as well as the accuracy and robustness of hyperpolarized 13 C imaging.
Related Concept Videos
Real Time RT-PCR
The real-time quantification of the number of amplified products is...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
IR Frequency Region: Fingerprint Region
Calibration Curves: Correlation Coefficient
Plotting and Calibrating the Root Locus
The maximum gain occurs at the breakaway points between open-loop poles on the real axis, while the minimum gain is...