Reliability of upper-limb diaphyseal mineral and soft-tissue measurements using peripheral Quantitative Computed
Mark A Jenkins1, Nicolas H Hart, Timo Rantalainen
1Centre for Exercise and Sports Science Research, Edith Cowan University, Perth, Australia.
Objectives:
To quantify between-day reliability of upper-body diaphyseal measurements (radius, ulna, humerus) using peripheral Quantitative Computed Tomography (pQCT).
Methods:
Fourteen males (age: 25.8±2.3 years,) underwent repeat pQCT scans (one to two days apart) at mid-shaft ulna (60%), mid-shaft radius (60%) and mid-shaft humerus (50%) cross-sections of the non-dominant limb. Intraclass correlation coefficients (ICC) and coefficients of variation (CV) were determined for musculoskeletal morphology variables.
Results:
Reliability was excellent (ICC: 0.76-0.99; CV: 1.3-7.3) at all sites for bone mass, stress-strain index, endocortical and pericortical radius, endocortical volumetric bone mineral density (vBMD), muscle area, total area, non-cortical area, and cortical area. Reliability was good to excellent (ICC: 0.58-0.80; CV: 0.6-3.7) for polar vBMD and mid-cortical vBMD; fair to excellent (ICC: 0.30-0.88; CV: 0.5-8.0) for muscle density and cortical density; and fair to good (ICC: 0.25-0.60; CV: 3.4-7.6) for pericortical vBMD. Average reliability across the three sites was excellent (ICC ≥0.77; CV ≤8.0).
Conclusions:
Overall between-day reliability of pQCT was excellent for the mid-shaft ulna, radius and humerus. pQCT provides a reliable and feasible body composition and skeletal morphology assessment tool for upper limb longitudinal investigations in scientific and clinic settings.
Related Concept Videos
Arteries of the Upper Limbs
Veins of Upper Limbs
The deep venous system is primarily composed of the ulnar and radial veins. The ulnar vein, which drains the fingers through the superficial palmar venous arches, and the radial vein, which serves the palms via the deep palmar...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Bones of the Upper Limb: Humerus
Bones of the Upper Limb: Ulna
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a...