Repeatability and reproducibility of variable flip angle T1 quantification in the prostate at 3 T
Xinran Zhong1,2, Sepideh Shakeri2, Dapeng Liu1
1Department of Radiological Sciences, David Geffen School of Medicine, University of California, California, Los Angeles, USA.
B1+ correction improves prostate T1 relaxation estimation repeatability and reproducibility. This enhanced accuracy in T1 mapping can lead to better quantification in dynamic contrast-enhanced MRI studies.
Area of Science:
- Magnetic Resonance Imaging (MRI)
- Quantitative MRI
- Prostate Imaging
Background:
- Variable Flip Angle (VFA) imaging is a common method for estimating T1 relaxation in the prostate.
- However, VFA imaging's sensitivity to B1+ inhomogeneity can limit its repeatability and reproducibility.
Purpose of the Study:
- To evaluate the repeatability and reproducibility of prostate T1 estimation.
- To compare T1 estimation with and without B1+ correction methods.
Main Methods:
- Prospective study involving 21 volunteers scanned twice on two 3T MRI scanners.
- Two B1+ mapping techniques (RR-VFA and satTFL) were used for correction.
- Intrascanner repeatability and interscanner reproducibility of T1 maps were assessed using concordance correlation coefficients and Bland-Altman plots.
Main Results:
- Both RR-VFA-corrected T1 and satTFL-corrected T1 demonstrated higher intra- and interscanner correlation compared to uncorrected VFA T1.
- Bland-Altman analysis revealed narrower limits of agreement and a smaller T1 range for corrected T1 values.
Conclusions:
- Applying B1+ correction (RR-VFA or satTFL) to VFA T1 significantly improves T1 estimation repeatability and reproducibility.
- This correction offers potential for enhanced quantification of prostate dynamic contrast-enhanced MRI parameters.
More Related Videos
09:00Experimental Methods for Spin- and Angle-Resolved Photoemission Spectroscopy Combined with Polarization-Variable Laser
Published on: June 28, 2018
09:14Nanotopology of Cell Adhesion upon Variable-Angle Total Internal Reflection Fluorescence Microscopy VA-TIRFM
Published on: October 2, 2012
Related Concept Videos
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Random Variables
Uppercase letters such as X or Y denote a random variable. Lowercase letters like x or y denote the value of a random variable. If X is a random variable, then X is written in words, and x is given as a number.
For example, let X = the...
Graphs of Equations in Two Variables
Variables Affecting Phosphorescence and Fluorescence
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...