Optimal combination of electrodes and conductive gels for brain electrical impedance tomography
Lin Yang1, Heng Li2, Junjie Ding3
1Department of Aerospace Medicine, Fourth Military Medical University, Xi'an, China.
Background:
Electrical impedance tomography (EIT) is an emerging imaging technology that has been used to monitor brain injury and detect acute stroke. The time and frequency properties of electrode-skin contact impedance are important for brain EIT because brain EIT measurement is performed over a long period when used to monitor brain injury, and is carried out across a wide range of frequencies when used to detect stroke. To our knowledge, no study has simultaneously investigated the time and frequency properties of both electrode and conductive gel for brain EIT.
Methods:
In this study, the contact impedance of 16 combinations consisting of 4 kinds of clinical electrode and five types of commonly used conductive gel was measured on ten volunteers' scalp for a period of 1 h at frequencies from 100 Hz to 1 MHz using the two-electrode method. And then the performance of each combination was systematically evaluated in terms of the magnitude of contact impedance, and changes in contact impedance with time and frequency.
Results:
Results showed that combination of Ag+/Ag+Cl- powder electrode and low viscosity conductive gel performed best overall (Ten 20® in this study); it had a relatively low magnitude of contact impedance and superior performance regarding contact impedance with time (p < 0.05) and frequency (p < 0.05).
Conclusions:
Experimental results indicates that the combination of Ag+/Ag+Cl- powder electrode and low viscosity conductive gel may be the best choice for brain EIT.
Related Concept Videos
Impedance Combination
Electrical Conductivity
In a practical conductor, an applied electric field may be sustained, causing a flow of electrons, which produce a current. The differential form of the current, the current density, is related to the electric field.
More generally, it is related to the force per unit charge, which involves the...
Electric Field of Parallel Conducting Plates
Consider a cross-section of a thin, infinite conducting plate having a positive charge. For such a large thin plate, as the thickness of the plate tends to zero, the positive charges lie on the plate's two large faces. Without an external electric field, the...
Standard Electrode Potentials
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...