Evaluation of Capillary and Other Vessel Contribution to Macular Perfusion Density Measured with Optical Coherence Tomography Angiography
Published on: February 18, 2022
CHOROIDAL NEOVASCULAR AREA AND VESSEL DENSITY COMPARISON BETWEEN TWO SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY
Avi Ohayon1, Riccardo Sacconi2,3, Oudy Semoun1
1Department of Ophthalmology, University Paris Est Creteil, Centre Hospitalier Intercommunal de Creteil, Creteil, France.
Purpose:
To compare choroidal neovascularization (CNV) area and vessel density (VD) measurements between two different swept-source optical coherence tomography angiography (SS-OCTA) devices.
Methods:
En face optical coherence tomography angiography (OCTA) images of patients affected by neovascular age-related macular degeneration were collected prospectively from two devices: Zeiss PLEX Elite 9000 (Carl Zeiss Meditec, Dublin, CA) and Topcon DRI OCT Triton SS-OCTA (Topcon, Tokyo, Japan). Choroidal neovascularization area and VD of images were measured and analyzed with ImageJ software by two readers to evaluate the agreement between two devices, with respect to different image size (3 × 3 and 6 × 6 mm) and different image segmentation (automatic vs. manual), and a Topcon equivalent Zeiss segmentation as control (i.e., the equivalent anatomical slab given by Topcon device on the Zeiss device).
Results:
A total of 30 eyes (30 patients) were analyzed. There was an excellent agreement between the two readers in CNV area measurements intraclass correlation coefficient >0.9 in all analyses. We found excellent agreement in CNV area measurements (manual and automatic segmentations) when comparing 3 × 3-mm or 6 × 6-mm images both for each single device and between the two devices (overall intraclass correlation coefficient > 0.9). Vessel density measurements between manual to automatic segmentation within the same device and same image size had a high intraclass correlation coefficient value, but there was a poor agreement in VD between different image sizes (3 × 3 mm vs. 6 × 6 mm) in the same device and also comparing the two devices (3 × 3 Topcon vs. 3 × 3 Zeiss; 6 × 6 Topcon vs. 6 × 6 Zeiss). There was a poor agreement between the Topcon equivalent Zeiss segmentation and all other segmentations.
Conclusion:
There was an excellent agreement in CNV area measurements for both swept-source optical coherence tomography angiography devices in automatic and manual segmentations. However, the Topcon equivalent Zeiss segmentation was not comparable with any of the preset segmentations of Topcon and Zeiss devices. There was a poor agreement in CNV VD between different image size and different devices. For these reasons, it seems that, for accurate longitudinal analysis of VD, it is better to use the same device for each individual, even if both devices can be used interchangeably for CNV area measurements using automatic or manual segmentations.
More Related Videos
03:47Retinal and Choroidal Thickness Changes in Populations with Helicobacter pylori Infection by Swept-Source Optical Coherence Tomography
Published on: November 1, 2024
09:56In Vivo Multimodal Imaging and Analysis of Mouse Laser-Induced Choroidal Neovascularization Model
Published on: January 21, 2018
Related Concept Videos
Sources of Self-Esteem III: Social Comparison
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Density
Molecular Comparison of Gases, Liquids, and Solids
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Current Density