Experimental Assessment of Proton Dose Calculation Accuracy in Small-Field Delivery Using a Mevion S250 Proton
Kyle D DePew1, Salahuddin Ahmad1, Hosang Jin1
1Department of Radiation Oncology, University of Oklahoma Health Sciences Centre, Oklahoma City, Oklahoma, USA.
Purpose:
Dose calculation accuracy of the Varian Eclipse treatment planning system (TPS) is empirically assessed for small-aperture fields using a Mevion S250 double scattering proton therapy system.
Materials And Methods:
Five spherical pseudotumors were modeled in a RANDO head phantom. Plans were generated for the targets with apertures of 1, 2, 3, 4, or 5 cm diameter using one, two, and three beams. Depth-dose curves and lateral profiles of the beams were taken with the planned blocks and compared to Eclipse calculations. Dose distributions measured with EBT3 films in the phantom were also compared to Eclipse calculations. Film quenching effect was simulated and considered.
Results:
Depth-dose scans in water showed a range pullback (up to 2.0 mm), modulation widening (up to 9.5 mm), and dose escalation in proximal end and sub-peak region (up to 15.5%) when compared to the Eclipse calculations for small fields. Measured full width at half maximum and penumbrae for lateral profiles differed <1.0 mm from calculations for most comparisons. In the phantom study, Eclipse TPS underestimated sub-peak dose. Gamma passing rates improved with each beam added to the plans. Greater range pullback and modulation degradation versus water scans were observed due to film quenching, which became more noticeable as target size increased.
Conclusions:
Eclipse TPS generates acceptable target coverage for small targets with carefully arranged multiple beams despite relatively large dose discrepancy for each beam. Surface doses higher than Eclipse calculations can be mitigated with multiple beams. When using EBT3 film, the quenching effect should be considered.
More Related Videos
07:38Functional Characterization of Na+/H+ Exchangers of Intracellular Compartments Using Proton-killing Selection to Express Them at the Plasma Membrane
Published on: March 30, 2015
10:03Proton Transfer and Protein Conformation Dynamics in Photosensitive Proteins by Time-resolved Step-scan Fourier-transform Infrared Spectroscopy
Published on: June 27, 2014
Related Concept Videos
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
¹H NMR of Labile Protons: Temporal Resolution
The –OH proton in alcohols typically appears in the range of δ 2 to 5 ppm but can vary depending on the specific...
¹H NMR of Labile Protons: Deuterium (²H) Substitution
¹H NMR Chemical Shift Equivalence: Homotopic and Heterotopic Protons
¹H NMR Chemical Shift Equivalence: Enantiotopic and Diastereotopic Protons
In chiral compounds such as 2-butanol, replacing the methylene hydrogens at C3 produces a pair of...
Acid Suppressive Drugs for Peptic Ulcer Disease: Proton Pump Inhibitors
Gastric acid, a potent cocktail of hydrogen and chloride ions, is produced in specialized parietal cells within the...