A Novel Technique for Raman Analysis of Highly Radioactive Samples Using Any Standard Micro-Raman Spectrometer
Published on: April 12, 2017
Alignment of the aberration-free XUV Raman spectrometer at FLASH
Mykola Biednov1, Günter Brenner2, Benjamin Dicke1
1Institut für Nanostruktur- und Festkörperphysik, Universität Hamburg, Notkestrasse 85, Hamburg, 22607, Germany.
Abstract:
An extreme-ultraviolet (XUV) double-stage Raman spectrometer is permanently installed as an experimental end-station at the PG1 beamline of the soft X-ray/XUV free-electron laser in Hamburg, FLASH. The monochromator stages are designed according to the Czerny-Turner optical scheme, adapted for the XUV photon energy range, with optical elements installed at grazing-incidence angles. Such an optical scheme along with the usage of off-axis parabolic mirrors for light collimation and focusing allows for aberration-free spectral imaging on the optical axis. Combining the two monochromators in additive dispersion mode allows for reaching high resolution and superior stray light rejection, but puts high demands on the quality of the optical alignment. In order to align the instrument with the highest precision and to quantitatively characterize the instrument performance and thus the quality of the alignment, optical laser interferometry, Hartmann-Shack wavefront-sensing measurements as well as off-line soft X-ray measurements and extensive optical simulations were conducted. In this paper the concept of the alignment scheme and the procedure of the internal optical alignment are presented. Furthermore, results on the imaging quality and resolution of the first monochromator stage are shown.
Related Concept Videos
IR Spectrometers
Mass Spectrometers
NMR Spectrometers: Overview
UV–Vis Spectrometers
Cavity Drainage and Flashings in Masonry walls
Weep holes, strategically placed at the base of the cavity, are critical for draining accumulated water. These openings are created by leaving head...
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and...